3M 245 Food Warmer User Manual


 
3M
TM
Ranger
TM
Blood/Fluid and Irrigation Warming Systems
English
12
Check the 3M
TM
Ranger
TM
system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A
Blood/Fluid Warming System
Preventative Maintenance Checklist
caution
Temperature adjustments cannot be made to the Ranger warming unit. If the test shows
the temperature is out of specication, please call 3M Patient Warming technical support at
1-800-733-7775.
toolS and equipment
• ermocouple: 0.005" lead or smaller (part #90020)
• Calibrated thermocouple temperature monitor
• Aluminum tape
• Ranger standard ow blood/uid warming set
• Phillips screwdriver
• Saline/water
0.75"
1.80"
Thermocouple
Inlet tube
Aluminum tape
Drain 2mL or more
from cassette to remove.
Insert cassette
before priming.
Fig. 1: Ranger standard ow blood/uid warming cassette
Model 245