Check the 3M
TM
Ranger
TM
system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524ACheck the 3M
TM
Ranger
TM
system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A
1. Use aluminum tape to ax a thermocouple
to the top surface of the Ranger warming as
shown in Figure 1.
2. Slide the cassette into the slot of the
Ranger warming unit.
3. Using room temperature saline/water,
prime the cassette per instructions
included with it.
note: To achieve correct results, uids
must be administered into the inlet tube,
NOT into the outlet tube.
4. Close patient line with white clamp to
stop ow.
5. Turn the Ranger warming unit ON and
allow it to warm up for 5 minutes.
6. e Ranger warming unit display should
read 41°C±1. Note the temperature on the
thermocouple temperature monitor. If the
temperature reading is not 41°C±1, call
3M Patient Warming technical support.
note: Most handheld temperature meters
have a ±1°C tolerance.
Fig. 2: Ranger standard ow blood/uid warming set
Blue
clamp
Fluid
source
White
clamp
Roller
clamp
Bubble
trap
Exit
Outlet
tube
Inlet
tube
Cassette
Model 245