These specifications apply to the Agilent
Technologies 8591E, 8593E, 8594E, 8595E, and
8596E spectrum analyzers.
Specifications
All specifications apply over O°C to +55°C. The
analyzer will meet its specifications after 2 hours
of storage at a constant temperature, within the
operating temperature range, 30 minutes after the
analyzer is tumed on, and after CAL FREQ and
CAL AMPTD (and for the 8593E, 8595E, and
8596E CAL YTF) have been run.
Frequency Specifications
Frequency Range
8591E
50 9 kHz to 1.8 GHz
75 1 MHz to 1.8 GHz
8593E 9 kHz to 22 GHz
Option 026/027 9 kHz to 26.5 GHz
Band LO harmonic = N
0 1 9 kHz to 2.9 GHz
1 1 2.75 GHz to 6.5 GHz
2 2 6.0 GHz to 12.8 GHz
3 3 12.4 GHz to 19.4 GHz
4 4 19.1 GHz to 22.0 GHz
4 4 (Opt. 026/027) 19.1 GHz to 26.5 GHz
8594E
dc coupled 9 kHz to 2.9 GHz
ac coupled 100 kHz to 2.9 GHz
8595E
dc coupled 9 kHz to 6.5 GHz
ac coupled 100 kHz to 6.5 GHz
8596E
dc coupled 9 kHz to 12.8 GHz
ac coupled 100 kHz to 12.8 GHz
Band LO harmonic = N
0 1 9 kHz to 2.9 GHz (dc coupled)
0 1 100 kHz to 2.9 GHz (ac coupled)
1 1 2.75 GHz to 6.5 GHz
2 2 6.0 GHz to 12.8 GHz
Frequency Reference (Opt. 004)
Aging ±2 x 10
-6
/year ±1 x 10
-7
/year
Temperature Stability ±5 x 10
-6
±1 x 10
-8
Initial Achievable Accuracy ±0.5 x 10
-6
±2.2 x 10
-8
Agilent 8590 E-Series
Portable Spectrum Analyzers
Data Sheet