Fisher & Paykel MR850 Cooktop User Manual


 
SPECIFICATIONS
Dimensions: 140 mm x 173 mm x 135 mm (without chamber fitted)
Weight: 2.8 kg (no chamber fitted), 3.1 kg (chamber fitted and filled with water)
Supply Frequency: 50/60 Hz Supply Voltage: 115 V~ Supply Current: 2.0 A max at 115 V~
Heater Plate: 150 W Heater Wire: 22 V~, 2.73 A, 60 W, 50/60 Hz
Heater Plate Over-temperature Cutout: 118 ± 6 °C
Temperature Control Settings
Invasive Mode: Chamber outlet: 35.5 - 37 °C, Airway 35 - 40 °C
Noninvasive Mode: Chamber outlet: 31 °C, Airway 28 - 34 °C
Display: Three digit 14 mm 7 segment LED
Range: 10 - 70 °C, Accuracy: ± 0.3 °C (in 25 - 45 °C temperature range)
Alarm Parameters
High Humidity Alarm: An immediate, audible alarm at a displayed temperature of 41 °C or if the
airway temperature exceeds 43 °C
Low Humidity Alarm: An audible alarm between 10 minutes @ 29.5 °C, and 60 minutes @ 34.5 °C
(Invasive Mode only)
Sound Pressure Level: Alarms exceed 50 dBA @ 1 m
Performance
Recommended ambient temperature range: 18 - 26 °C
CAUTION:
If operating in ambient temperatures outside the recommended range, consult your local
Fisher & Paykel Healthcare representative.
Recommended Flow Range: Invasive Mode: up to 60 L/min, Noninvasive Mode: up to 120 L/min
Refer to breathing circuit specifications for minimum flow
Humidity Performance: Invasive Mode: > 33 mg/L, Noninvasive Mode: >10 mg/L
Maximum Operating Pressure: Refer to chamber and breathing circuit specifications
Warm-up time: Less than 30 minutes
Symbol Definitions
Type BF ATTENTION
Consult
accompanying
documents
Alternating
Current
Drip
Proof
CAUTION
Hot surfaces
may exceed
85 °C
Respiratory Humidifier Classified by
Underwriters Laboratories Inc. with respect to
electric shock, fire and mechanical hazards
only in accordance with UL 60601-1 and
CAN/CSA C22.2 No. 601.1 84DJ
CLEANING
MR850 Heaterbase Using a damp cloth, clean the humidifier with either of the following: Isopropyl Alcohol, normal dishwashing detergent. Temperature
Probe Always clean before use. Probes can be cleaned with one of the following solutions: Sporicidin
TM
, Cidex
TM
OPA, Sporox
TM
OR sterilize the probe using Ethylene Oxide
sterilization at 55 °C (131 °F), 80 kPa, (allow at least 15 hours for residual ETO to disperse before use). Wipe the airway temperature probes clear of any cleaning residues
before use. Store probes in clean conditions. DO NOT autoclave probes.
CAUTION:
DO NOT immerse the heaterbase or temperature probe electrical connections in any liquid.
NOTE: Follow the cleaning agent manufacturers’ instructions carefully. It is the user’s responsibility to qualify any deviations from these procedures, both for disinfecting
efficacy and physical effect on the probe and the heaterbase. See the MR850 technical manual for more detailed instructions on cleaning.
WARNING
Ensure that invasive mode is set for patients
that have bypassed airways.
The use of breathing circuits, chambers or
other accessories which are not approved
by Fisher & Paykel Healthcare may impair
performance or compromise safety.
Ensure that both temperature probe sensors
are correctly and securely fitted. Failure to
do so may result in temperatures in excess
of 41 °C being delivered to the patient.
Ensure maintenance of grounding integrity
by connection to a "hospital grade"
receptacle. Always disconnect supply before
servicing.
When mounting a humidifier adjacent to a
patient ensure that the humidifier is always
positioned lower than the patient.
The operation of high frequency surgical
apparatus, shortwave or microwave
equipment in the vicinity of the humidifier
may adversely affect its function. If this
occurs, the humidifier should be removed
from the vicinity of such devices.
Do not touch the glass tip of the chamber
temperature probe during use. Keep black
connectors dry at all times.
Visually inspect accessories for
damage before use.
Patents Pending in Principal Countries
of the World
INSTRUCTION SHEET
DO NOT DISCARD
REF 185042343
Rev G 2004-02
MR850 RESPIRATORY HUMIDIFIER
The MR850 respiratory
humidifier is used to
warm and humidify
gases delivered to
patients requiring
mechanical ventilation,
positive pressure
breathing assistance,
or other medical gases.