Fisher & Paykel MR850 Cooktop User Manual


 
NOTE
To set the humidifier to non-heater wire mode, refer to the MR850
technical manual. (Rev. H or later).
If a heated breathing circuit is used while in non-heater wire mode
the humidifier automatically changes to heater wire mode. To use
non-heater wire mode again will require it to be manually set.
The airway probe out test does not work with non-heated breathing
circuits. A low humidity alarm will instead be given.
When there is no gas flow the humidifier will give a low humidity
alarm.
If the chamber runs dry either a water out alarm or a low humidity
alarm will occur.
(a)
(c)
(d)
(e)
(b)
NON-HEATED BREATHING CIRCUITS
SET UP
Slide humidification chamber (a) onto humidifier base and
connect breathing circuit (b) (refer to humidification chamber and
breathing circuit operating instructions for further details).
Connect the temperature probe plug (c) (REF 900MR86X) to the
blue socket on the humidifier base until an audible click is heard.
Push the chamber probe (d) and airway probe (e) into the
breathing circuit. Make sure the chamber probe is correctly
located in its key-way and that both probes are pushed home.
The probe lead can be restrained using breathing circuit clips.
The humidification system is now set up and ready for use. After
power on, the humidifier will default to invasive mode.
RECOMMENDED SETUP FOR ADULT PATIENTS
Non-heated breathing circuit with MR290 Chamber
900MR869 Temperature Probe, Mounting bracket to suit ventilator
For all other applications refer to Product Guide or contact your local
Fisher & Paykel Healthcare representative.
SPECIFICATIONS (additional)
Temperature Control Settings (non heated circuits)
Invasive Mode: Airway 37 ºC
Noninvasive Mode: Airway 31 ºC
Alarm Parameters (non heated circuits)
High Temperature Alarm: Immediate, audible alarm at a displayed
temperature > 41 ºC
Low Temperature Alarm: Audible alarm at a displayed temperature
< 29 ºC (Invasive), < 26 ºC (Noninvasive)
Performance (non heated circuits)
Recommended Flow Range: Invasive Mode: 6 to 60 L/min
Non-Invasive Mode: 6 to 120 L/min
A
TTENTION: Refer to operating instructions which
accompany each accessory.