Function and Instruction For 2-8 Pharmaceutical
Refrigerator
5Adjust alarmtemperature range:See fig #1
6 Defectsdisplay
7Inner light
Two methodsset high temperaturealarm “HA”and lowtemperature alarm “LA”
Press “Alarm” buttonfor 10 seconds. LED displaycode “HA”,press “SET” button again(quickly),code
“HA” flash, adjust the set- temperature by both pressing “+” and ” buttons and increment 1°C,
adjustment range shouldn't be 6°C higher than pre-set cabinet temperature,press “Alarm” button to
return the former menu(each press of “Alarm” button will return one step back to former menu until
display cabinet temperature) and LED display “HA” Press both “+” and ” to choose another
parameter “LA”, “LA” setting method like “HA”, and its adjustment range shouldn't be 6° C lower than
pre-set cabinettemperature.
The control panel only can display one sensor error, when the sensor error happens.LED display
“E1”,and alarm at the same time. The refrigerator will keep working according to the defect safty
model,and you should put the stored goods to other refrigerator as soon as possible,and cantact the
after-sales departmentquickly.
In orderto see thestored goods clearly,a LED light hasbeen installed inthe cabinet.You cancontrol
light “on” or “off”by “light” button to save energe,please turn off the light after observing the stored
goods inthe refrigerator.
The lightis onwhen thedoor is open;The lightis off when thedoor isclosed.