GENERAL INFORMATION
INTRODUCTION
The REMCOR CH Series Recirculating Liquid Chiller is designed to provide an accurate, reliable, and user--
friendly system for cooling a continuous flow of pure liquid and keep that liquid at a constant temperature in vari-
ous closed loop or tank cooling applications.
The CH Series Chiller consists of an air--cooled refrigeration system housed in a sturdy sheet metal frame and
cabinet. A standard pump and insulated liquid reservoir package provides a complete liquid cooling and circu-
lating system.
The CH Series Chiller is designed to operate in a clean laboratory or industrial environment where ambient
temperatures range from 40 to100° F (5 to 38° C). With proper installation, operation, and maintenance, the
CH Series Chiller will provide years of trouble free service.
UNPACKING AND INSPECTION
This unit was thoroughly inspected before leaving the factory and the carrier has accepted and signed for it.
Any damage or irregularities should be noted at the time of delivery and immediately reported to the carrier. Re-
quest a written inspection report from the Claims Inspector to substantiate any necessary claims. In the event
that an immediate replacement is necessary, please contact REMCOR Chiller Sales at 1--800--551--4423.
DESIGN DATA
Table 1. Design Data
CH951 CH1500 CH1502 CH1503
Cooling Capacity:
BTU/hr (W) at 80_ F(27_ C) and
70_ F (21_ C) Liquid Temperature.
12,000 (3,515) 18,000 (5,272) 18,000 (5,272) 18,000 (5,272)
Compressor Horsepower 1 (.746 kW) 1 1/2 (1.12 kW) 1 1/2 (1.12 kW) 1 1/2 (1.12 kW)
Electrical Data:
Voltage/Phase/Hertz/Amperage
230/1/60
11.0 Amps
230/1/60
13.0 Amps
230/3/60
16.4 Amps
460/3/60
10.0 Amps
Refrigerant Type: R134a R22 R22 R22
Reservoir Capacity 6.0 gallons (22.7
liters)
6.0 gallons (22.7
liters)
6.0 gallons (22.7
liters)
6.0 gallons (22.7
liters)
Physical Dimensions, Width ¢ Depth
¢ Height:
22I¢ 26.5XI¢
38.25I
(56cm ¢ 67cm
¢ 97cm)
22I¢ 26.5XI¢
38.25I
(56cm ¢ 67cm
¢ 97cm)
22I¢ 26.5XI¢
38.25I
(56cm ¢ 67cm
¢ 97cm)
22I¢ 26.5XI¢
38.25I
(56cm ¢ 67cm
¢ 97cm)
Fittings:
Process Connections 3/4 FPT (S/S) 3/4I FPT (S/S) 3/4I FPT (S/S) 3/4I FPT (S/S)
CHILLER INSTALLATION
Location of Chiller
THE CHILLER MUST BE LOCATED NEAR A PROPERLY GROUNDED ELECTRICAL OUTLET. THE CIR-
CUIT SHOULD BE FUSED AND NO OTHER ELECTRICAL APPLIANCE SHOULD BE CONNECTED TO THE
CIRCUIT. ALL ELECTRICAL WIRING MUST CONFORM TO NATIONAL AND LOCAL ELECTRICAL
CODES.
The Chiller must be located in a well ventilated, indoor area where ambient temperatures will remain above
40_ F (5_ C) and will never increase above 100_ F (38_ C). To obtain optimum cooling capacity, the ambient
temperature should be at or below 80_ F (27_ C).