their respective owners.
Model No.
A-Depth
in. (mm)
B-Width
in. (mm)
C-Height
in. (mm)
Weight
lbs. (kg)
Voltage
(60 Hz)
Voltage
(50 Hz)
PS018A 36 (914) 40 (1016) 25.25 (641)) 394 (178) 208/230, 1-phase 200/230, 1-phase
PS024A 36 (914) 40 (1016) 25.25 (641) 404 (183) 208/230, 1-phase 200/230, 1-phase
PS036A 36 (914) 48 (1219) 32.5 (825) 515 (233) 460, 3-phase
208/230, 1- or 3-phase
400, 3-phase
200/230, 1- or 3-phase
PS048A 36 (914) 53 (1346) 38.5 (978) 592 (268) 460, 3-phase
208/230, 3-phase
400, 3-phase
200/230, 1- or 3-phase
PS060A 36 (914) 53 (1346) 38.5 (978) 642 (291) 460, 3-phase
208/230, 3-phase
400, 3-phase
200/230, 1- or 3-phase
ACCESS, AND AIR FLOW.
FOR COMPONENT REMOVAL,
RECOMMENDED CLEARANCE
INDICATE A
SHADED AREAS
H
H
I
I
(25mm)
1" MAX.
(25mm)
1" MAX.
(914mm)
36"
(914mm)
36"
UNIT DIMENSIONAL DATA
OUT
AIRFLOW
B
IN
AIRFLOW
C*
D
A
1" NPT
(914mm)
36"
(51mm)
2"
(457mm)
18"
ACCESS, AND AIR FLOW.
FOR COMPONENT REMOVAL,
RECOMMENDED CLEARANCE
INDICATE A
SHADED AREAS
PRESSURE GAUGES
OPTIONAL REFRIGERANT
UNIT DIMENSIONAL DATA
Standard Features
• Indoor/outdoor cabinet
• Hermetic compressor
• Refrigerant dehydrator
• Expansion valve
• Refrigerant receiver
• Stainless steel fluid reservoir expansion tank
• Pressure relief valve
• Hot gas bypass valve capacity control
Optional Features
• Refrigerant gauges and reservoir heater
• Stainless steel pump
Accessories
• Emergency water bypass
• Enviro-source emergency bypass
• Water level control and alarm system
•Fluid pressure gauge
•Check valve
• Dial thermometer
• Relief valve
• Circuit setter
• Ball valve
• High temperature thermostat
• Low flow switch
• Liebert RCM8 monitor