ResMed 150 Ventilation Hood User Manual


 
Global leaders in sleep and respiratory medicine www.resmed.com
Stellar 150 system and accessories—ordering information
Stellar 150 features
H4i Heated Humidifier
for Stellar
ANZ 26951
APAC 24930
Stellar 150
ANZ and APAC 24162
HumiCare D900
18405
Mobility bag
24977
Pulse oximeter
22305 (Oximeter cable)
16936 (1 sensor and
25 tapes)
FiO
2
monitoring kit
(does not include FiO
2
monitoring sensor
CAP008617) 24955
Remote Alarm II
(does not include cables)
ROW 27902
ResMed Power Station II
Battery and bag 24921
DC cable 24959
Power supply 90W
ANZ and APAC 36824
ResMed leak valve
24988
Distributed by: ResMed Ltd Bella Vista, NSW, Australia +61 (2) 8884 1000 or 1 800 658 189 (toll free). ResMed Corp 9001 Spectrum Center Boulevard San Diego CA 92123 USA.
See www.resmed.com for other ResMed locations worldwide. H4i, ResScan, SlimLine, Stellar and TiControl are trademarks of ResMed Ltd. Stellar is registered in U.S. Patent
and Trademark Office. © 2012 ResMed Ltd. Specifications may change without notice. 1014183/4 2012-09
ResMed Germany Inc.
Fraunhoferstr. 16 82152
Martinsried Germany.
RESMED POWER STATION II
BATTERY TECHNOLOGY Lithium-ion
RECHARGE TIME 4 hours from fully empty to
fully charged
BATTERY DURATION 8 hours per battery at
EPAP 5 cm H
2
O, IPAP
15 cm H
2
O and 20 bpm (two
batteries can be connected
simultaneously. See
www.resmed.com/RPSII
for more details)
CAPACITY <100 Wh
DIMENSIONS 230 x 126 x 26 mm
(L x W x H)
BATTERY WEIGHT 0.9 kg
POWER SUPPLY Input range 100–240 V,
50–60 Hz, 1.0-1.5 A
Nominal for aircraft use
110 V, 400 Hz
For more information
on Stellar 150
please visit
www.stellar150.com
STELLAR 150
MODES iVAPS (intelligent Volume-Assured
Pressure Support)
S/T (Spontaneous/Timed)
PAC (Pressure Assisted Control)
CPAP (Continuous Positive
Airway Pressure)
S (Spontaneous)
T (Timed)
PRESSURE RANGE IPAP: 2–40 cm H
2
O
EPAP: 2–25 cm H
2
O
TiCONTROL Ti Max 0.3–4 sec
Ti Min 0.1–Ti Max
RESPIRATORY RATE 5–60 bpm
RISE TIME Min, 150–900 msec (approx.)
TRIGGER AND CYCLE Five sensitivity settings
ADJUSTABLE High Leak
ALARMS Low Minute Ventilation
High Pressure
Low Pressure
High Respiratory Rate
Low Respiratory Rate
Apnea
High FiO
2
Low FiO
2
Low SpO
2
Non-vented mask
STANDARD FIXED Circuit disconnected
ALARMS Over pressure
Blocked tube
Internal battery empty
(See Clinical Guide for
full list of alarms)
DIMENSIONS 230 x 170 x 120 mm
(L x W x H)
WEIGHT 2.1 kg
AIR FILTER Electrostatic fibre mesh
AIR OUTLET 22 mm taper, compatible with
ISO 5356–1:2004
INTERNAL BATTERY Lithium-ion, 2 hours at EPAP 5,
IPAP 15, 20 bpm
POWER SUPPLY AC 100–240V 50–60Hz,
2.2A AC 110V 400Hz,
2.2A DC 24V/3A
IEC 60601–1 Class II (double insulation)
CLASSIFICATION Type BF Continuous operation
Specifications