ELISEE 150 TECHNICAL SPECIFICATIONS Paediatric Adult Adjustments
PARAMETERS
Pressure slope 1, 2, 3, 4 1, 2, 3, 4
Inspiration pressure (cm H
2
O) 3–40 in NIV
5–60
1 cm H
2
O
3–60 in Invasive
Pressure support (cm H
2
O) 3–40 in NIV
5–60
1 cm H
2
O
3–60 in Invasive
PEEP (cm H
2
O) 0–20 0–25 1 cm H
2
O
CPAP (cm H
2
O) 3–20 3–25 1 cm H
2
O
Tidal volume (mL)
50–500
300–2500
10 mL (50–1000)
100 mL (1000–2500)
Security tidal volume (mL)
NO / 50–500
NO / 300–2500
10 mL (50–1000)
100 mL (1000–2500)
Maximum flow setting (volume) (L/min) 5–40 10–120 1 L/min
Flow shape (1–5) Square/Decreasing (1–5) Square/Decreasing
I/E ratio (1/x) 1/0.4–1/9.9 1/0.4–1/9.9 1/0.1
Respiratory rate 2–80 2–50 1 bpm
Minimum breath (bpm) NO / 3–80 NO / 6–50 1 bpm
Maximum inspiration time (seconds) 0.3–3.0 0.3–3.0 0.1 sec
Apnea time (seconds) 5–20 10–60 1 sec
TRIGGERS
Inspiration pressure trigger (cm H
2
O) 0.2–5.9 / NO in 0.2–5.9 / NO in
(Invasive ventilation only) (A)CV and P(A)CV modes (A)CV and P(A)CV modes
0.1 cm H
2
O
0.2–6.0 in PS.SV, PS.V
T
, 0.2–6.0 in PS.SV, PS.V
T
, 
SIMV and PSIMV modes SIMV and PSIMV modes
Inspiration flow trigger (L/min) 0.2–9.9 / NO in 0.2–9.9 / NO in
(Invasive ventilation only) (A)CV and P(A)CV modes (A)CV and P(A)CV modes
0.1 L/min
0.2–10.0 in PS.SV, PS.V
T
, 0.2–10.0 in PS.SV, PS.V
T,
SIMV and PSIMV modes SIMV and PSIMV modes
Inspiration trigger AUTO / 1–5 in PS.SV, AUTO / 1–5 in PS.SV,
(Noninvasive ventilation only) PS.V
T
, SIMV and PS.V
T
, SIMV and
PSIMV modesPSIMV modes 
AUTO / 1–5 / NO in AUTO / 1–5 / NO in
(A)CV and P(A)CV modes (A)CV and P(A)CV modes
Expiratory trigger 10–90 / AUTO in PS.SV, PS. 10–90 / AUTO in PS.SV, PS.
(% peak flow) V
T
, SIMV and PSIMV modes V
T
, SIMV and PSIMV modes
1 %
LUNG RECRUITMENT
Period (minutes) NO / 1–60 NO / 1–60
1 min
Duration (second
s)
0.2–40
0.3–40
0.1 sec (0.2–3)
1 sec (3–40)
Pressure (cm H
2
O) 5–60 5–60 1 cm H
2
O
TECHNICAL DATA
External DC supply 12–28 VDC, maxi 15 amps
Internal battery Li-Ion, 14.4 VDC, 6.3 AH
External battery Li-Ion, 14.4 VDC, 6.3 AH
Power consumption max 75 VA
External power 12–28 V
Internal battery capacity*  4–6 hrs
External battery capacity* 4–6 hrs
Size 26 cm x 24 cm x 13 cm (10.2” x 9.4” x 5.1”)
Mains supply 100 VAC/1 amp to 230 VAC / 0.6 amp – 50/60 Hz
Weight 4.45 kg (9.79 lb) with internal and external battery pack
* Tested under specific settings. Please refer to the User Guide.