Sanyo MDF-U32V Freezer User Manual


 
Specifications
*Appearance and specifications are subject to change without notice.
*Cooling performance is indicated by the temperature reached at the center of the freezer (at ambient temperature of 30°C with no load).
In order to use the freezer at a stable temperature for a long time, it is recommended that the temperature be set to at least 5°C higher
than the indicated lowest temperature. In addition, depending on the usage conditions, it may not be possible to reach the indicated
lowest temperature.
HFC
Model
LCO2 backup system
LN
2 backup system
Backup system mounting kit
Circular recorder
Chart paper
Ink pen
Slip chart recorder
Chart paper
Ink pen
Recorder mounting bracket
Sensor cover
Storage racks
3 drawers (for lower chamber)
SANYO DAQ (
Data Acquisition
)
system
MDF-C8V
CVK-UB4
MDF-UBK
MTR-G85
RP-G85
PG-R
MTR-85H
RP-85
DF-38FP
MDF-S3085
(for MTR-85H)
MTR-C8
MTR-480
MTR-2000
MDF-U32V
CVK-UB2
CVK-UBN2
MTR-G85
RP-G85
PG-R
MTR-85H
RP-85
DF-38FP
MDF-S3085
(for MTR-85H)
MDF-30SC
MDF-30R
MTR-480
MTR-2000
MDF-U53V
CVK-UB2
CVK-UBN2
MTR-G85
RP-G85
PG-R
MTR-85H
RP-85
DF-38FP
MDF-S3085
(for MTR-85H)
MDF-70SC
MDF-50R
MTR-480
MTR-2000
MDF-U73V
CVK-UB2
CVK-UBN2
MTR-G85
RP-G85
PG-R
MTR-85H
RP-85
DF-38FP
MDF-S3085
(for MTR-85H)
MDF-70SC
MTR-480
MTR-2000
MDF-C2156VAN
Equipped
MTR-155H
RP-155
DF-38FP
MDF-S30150
(for MTR-155H)
MTR-480
MTR-2000
Options
Distributed by:
We design and manufacture products in accordance with ISO9001 & 14001 requirements.
RoHS (European Restriction of Hazardous Substances directives) compliant
• All data shown here are measured based on our company standards. If detailed
data is required, please contact our company.
• If product inspection records are required, please contact our company. (There
will be a charge for the records.) Please note that the records do not include tem-
perature adjustment range or data on temperature distribution inside the unit.
Performance Data
40
20
0
–20
–120
–40
–60
–80
–100
076528341
Temperature (˚C) Temperature (˚C)
Time (hours)
525201510 30 35
20
10
15
5
Time (hours) Time (hours)
Ambient temperature (˚C)
Ambient temperature (˚C)
02010 30 40
20
25
10
15
5
Time (hours)
40
20
0
–20
–120
–40
–140
–60
–160
–80
–100
02 2046810 181612 14
Time (hours)
Time (hours)
Ambient temperature (˚C)
Temperature (˚C)
525201510 30 35
20
10
15
5
Ambient temp. and maintenance
of –70°C by injection of LCO
2 (30kg)
525201510 30 35
20
10
15
5
Time (hours)
Ambient temperature (˚C)
525201510 30 35
10
15
20
5
Time (hours)
Ambient temperature (˚C)
MDF-C2156VAN
LN2 backup system
operation characteristics
MDF-C8V
MDF-U32V
Central air temp. in cabinet
Ambient temp. 30°C, No-load
Ambient temp. and maintenance
of –70°C by injection of LCO
2 (30kg)
Central air temp. in cabinet
Ambient temp. 30°C, No-load
Ambient temp. and maintenance
of –70°C by injection of LCO
2 (30kg)
Central air temp. in cabinet
Ambient temp. 30°C, No-load
Ambient temp. and maintenance
of –135°C by injection of LN
2 (50kg)
Central air temp. in cabinet
Ambient temp. 30°C, No-load
Ambient temp. and maintenance
of –70°C by injection of LCO
2 (30kg)
Central air temp. in cabinet
Ambient temp. 30°C, No-load
MDF-U73V
MDF-U53V
Temperature (˚C)
Time (hours)
40
20
0
–20
–120
–40
–140
–60
–160
–80
–100
0126543210111987
40
20
0
–20
–120
–40
–140
–60
–160
–80
–100
0126543210111987
Time (hours)
Temperature (˚C)
40
20
0
–20
–120
–40
–140
–60
–160
–80
–100
0126543210111987
CO2 backup system
operation characteristics
Pull-up characteristics
during power failure
Pull-down characteristics
Further details regarding our products can be accessed at http://www.sanyo.co.jp/cmg/biomedical
Model
Temperature range
Maximum cooling performance
Exterior dimensions (W x D x H)
Interior dimensions (W x D x H)
Net weight
Effective capacity
Shelves
Access port
Compressor
Refrigerant
Alarm system
Remote alarm contact
Accessories
Model
Temperature range
Maximum cooling performance
Exterior dimensions (W x D x H)
Interior dimensions (W x D x H)
Net weight
Effective capacity
Access port
Compressor
Refrigerant
Alarm system
Remote alarm contact
Accessories
MDF-C8V
-60 to -80°C (1°C increments)
-80°C (Ambient temp. 30°C)
550 x 685 x 945 mm (21.6” x 27.0” x 37.2”)
405 x 490 x 425 mm (15.9” x 19.3” x 16.7”)
Approx. 67 kg (148lbs.)
84 Liters (3.0 cu.ft.)
17 mm diameter, 2 locations backside, bottom
Rotary type, Output: 400W
High/Low temp. alarm (±5°C to ±20°C),
Power failure alarm, Remote alarm contact:
DC30V, 2A
1 set of keys, 1 scraper
MDF-U73V
1010 x 875 x 2010mm
(39.8” x 34.4” x 79.1”)
870 x 600 x 1400mm
(34.2” x 23.6” x 55.1”)
Approx. 350kg (772 lbs.)
728 Liters (25.7 cu.ft.)
MDF-U53V
–50˚C to –86˚C (1˚C increments)
–86˚C (Ambient temp. 30˚C)
770 x 875 x 1990mm
(30.3” x 34.4” x 78.3”)
630 x 600 x 1380mm
(24.8” x 23.6” x 54.3”)
Approx. 299kg (659 lbs.)
519 Liters (18.3 cu.ft.)
Stainless steel, 3 shelves
17mm diameter, 3 locations (back, bottom left/right corners)
Hermetic type, Output:
1100 W (high stage side), 1100 W (low stage side)
High/low temperature, Power failure, Filter check,
Self diagnostics, Door check,
Part replacement nortification
Allowable contact capacity: DC 30V, 2A
1 set of keys, 1 scraper
MDF-U32V
670 x 867 x 1860mm
(26.4” x 34.1” x 73.2”)
490 x 600 x 1140mm
(19.3” x 23.6” x 44.9”)
Approx. 258kg (569 lbs.)
333 Liters (11.8 cu.ft.)
Hermetic type, Output:
400 W (high stage side),
600 W (low stage side)
Temperature, Power failure,
Filter check, Self diagnostics,
Remote alarm contact
MDF-C2156VAN
-125°C to -150°C (1°C increments)
-150°C (Ambient Temp. 30°C)
1730 x 765 x 1010 mm (68.1” x 30.1” x 39.8”)
760 x 495 x 615 mm (29.9” x 19.5” x 24.2”)
Approx. 319 kg (703lbs.)
231 Liters (8.2 cu.ft.)
40 mm diameter, 1 location
Hermetic type, Output: 1100W (high stage
side), 1100W (low stage side)
HFC
High/Low temperature, Power failure,
Filter check, Self diagnosis, Door check,
Part replacement notification
Allowable contact capacity, DC 30V, 2A
1 set of keys, 1 scraper, Connecting tube for
LN
2 backup system
SANYO Electric Co., Ltd.
Biomedical Division
©2008 SANYO Printed in Japan 2008.3 MA
SHR119