SANYO E&E America Company
1062 Thorndale Avenue, Bensenville, IL 60106 USA
Toll Free USA 800-858-8442 • Fax 630-238-0074
www.sanyobiomedical.com
SANYO Canada, Inc.
1- 300 Applewood Crescent, Concord, Ontario L4K 5C7
905-760-4025 • Fax 905-760-9945
*Appearance and specifications are subject to change without notice.
*Cooling performance is indicated by the temperature reached at the center of the freezer (at ambient temperature of 30°C with no load). In order to use the freezer at a
stable temperature for a long time, it is recommended that the temperature be set to at least 5°C higher than the indicated lowest temperature. In addition, depending on the
usage conditions, it may not be possible to reach the indicated lowest temperature.
■
The newly developed refrigeration system and freezer struc-
ture minimizes intrusive operating noise.
■
Achieves 10% higher energy conservation than competing
models.
■
The newly developed back cover combined with new aero-
dynamically designed and placed components in the refrigera-
tion compartment provide superior air flow, making it possible
to drastically reduce the stress to the freezer and contributing
to excellent durability.
■
Two independent and insulated inner doors ensure maxi-
mum interior chamber uniformity.
■
A new rugged style inner door latch tightly closes the inner
door and makes accessing chamber contents easier.
■
The condenser filter is located at the bottom right side of
the front panel to make filter removal and cleaning easier.
■
Alarm lamp and buzzer offer secure warning of power fail-
ure or abnormal temperature increase.
■
New rugged, one-handed outer door latch has a hasp to
allow a padlock for additional security
■
High and low temperature warning provides an audible and
visual alarm when the temperature deviates more than ±5°C
to ±20°C (adjustable) from the set point.
■
Alarm ring back function ensures that buzzer will resume
operation should alarm conditions continue after it is silenced.
■
Microprocessor-controlled filter-clog check function
protects the refrigeration circuit.
■
Control panel with digital display for easy operation.
EVOLUTIONARY DESIGN
ACCESSIBILITY
SAFETY
p r e s e r v a t i o n
w w w . s a n y o b i o m e d i c a l . c o m
MDF-U7386S
Dimensions [Unit : mm (inch)]
Specifications
Model
Temperature range
Maximum cooling
performance
Exterior dimensions
(W x D x H)
Interior dimensions
(W x D x H)
Net weight
Effective capacity
Shelf
Access port
Compressor
Refrigerant
Alarm
Remote alarm contact
Accessories
Power Requirement
MDF-U7386SC
–50°C to –86°C (1°C increment)
–86°C
(Ambient temp. 30°C)
1130 x 875 x 1990mm
(44.5” x 34.4” x 78.3”)
870 x 600 x 1280mm
(34.2” x 23.6” x 50.4”)
Approx. 355kg (783 lbs.)
668 L (23.5 cu.ft.)
Stainless steel, Adjustable,
3 shelves, W848 x D533mm,
Max. load: 50kg (110 lbs.)/shelf
17mm diameter, 3 locations
(back, bottom left/right corner)
Hermetic type, Output: 1100 W
(high stage), 1100 W (low stage)
HFC refrigerants
High/low temperature,
Power failure, Filter check,
Battery check
Allowable contact capacity:
DC 30V, 2A
1 set of keys, 1 scraper
220V / 60 Hz
Built-in voltage booster
*Appearance and specifications are subject to change without notice.
ISO 9001 & 14001 certified
Distributed by:
Performance Data
Options
Circular recorder
MTR-C954
Strip Chart Recorder
MTR-85H
Recorder mounting
bracket
MDF-S3085
(for MTR-85H)
LCO2 backup system
CVK-UB2 (I)
LN2 backup system
CVK-UBN2
40
20
0
–20
–40
–60
–80
–100
0 2 4 6 8 10
50
12
352520
14
1510
Temperature (°C)
40
20
0
–20
–40
–60
–80
–100
0 2 4 6 8 10 12 14
Temperature (°C)
Time (hour)
Time (hour)
Pull-up characteristics
during power failure
30
12
8
6
4
2
10
0
Time (hour)
Ambient temperature (°C)
CO2 backup system operation
characteristics
Central air temp. in cabinet
Ambient temp. 30°C, No-load
Ambient temp. and maintenance
of –70°C by injection of LCO2 (30kg)
Inventory rack
SUR-624-I
Rack Capacity: 24
SANYO Electric Co., Ltd.
SANYO Sales & Marketing Corporation
5-15, Hiyoshi-cho 2-chome, Moriguchi City, Osaka, 570-8634, Japan
Telephone: +81-6-6992-0322 Fax: +81-6-6992-9291
URL http://www.sanyo.co.jp/cmg/biomedical
©2006 SANYO Printed in Japan 2006.3. MA. SHR101
18
(0.7)
48
(1.9)
1130 (44.5)
870 (34.3)
1196 (47.1)
1540 (60.6)450 (17.7)
1990 (78.3)
605.5 (23.8)
105
(4.1)
1280 (50.4)
130 (5.1)130 (5.1)
140
(5.5)
795 (31.3)
867 (34.1)
11
(0.4)
1625 (64)
Features
EVOLUTIONARY DESIGN
• The newly developed refrigeration system
and the freezer structure offer a quiet
experimental environment.
• Achieves 10% higher energy conservation
than previous model. (For model rated 220V, 50Hz)
• The newly developed back cover combined
with new aerodynamically designed and
placed components in the refrigeration
compartment provide superior air flow,
making it possible
to drastically
reduce the stress
to the freezer and
contributing to
excellent durability.
• Two independent and insulated inner doors
prevent cold air leakage.
ACCESSIBILITY
• A new beak style inner
door latch tightly closes
the inner door against
the freezer frame. It
also helps make
opening and closing the door smoother.
• The condenser filter is situated at the
bottom right side of the front panel to make
filter removal and cleaning easier.
Back
cover
Compressors
Front
Rear
Air
SAFETY
• Alarm lamp and buzzer offer secure warning
of power failure or abnormal temperature
increase.
• High and low temperature warning provides
an audible and visual alarm when the
temperature deviates more than 5°C to
20°C (adjustable) from the set point.
• Alarm ring back function ensures that buzzer
will resume operation, should alarm
conditions continue after it is silenced.
• Microprocessor-controlled filter-clog check
function protects the refrigeration circuit.
• New rugged, one-handed
outer door latch has a hole to
allow a padlock to securely
protect valuable samples.
• Control panel with
digital display for easy
operation.
Pull-down characteristics
*Cooling performance is indicated by the temperature reached at
the center of the freezer (at ambient temperature of 30C with no
load). In order to use the freezer at a stable temperature for a long
time, it is recommended that the temperature be set to at least 5C
higher than the indicated lowest temperature.
In addition, depending on the usage conditions, it may not be
possible to reach the indicated lowest temperature.
MDF-U7386S
Dimensions
[Unit : mm (inch)]
Specifications
Model
Temperature range
Maximum cooling
performance
Exterior dimensions
(W x D x H)
Interior dimensions
(W x D x H)
Net weight
Effective capacity
Shelf
Access port
Compressor
Refrigerant
Alarm
Remote alarm contact
Accessories
Power Requirement
MDF-U7386SC
–50°C to –86°C (1°C increment)
–86°C
(Ambient temp. 30°C)
1130 x 875 x 1990mm
(44.5” x 34.4” x 78.3”)
870 x 600 x 1280mm
(34.2” x 23.6” x 50.4”)
Approx. 355kg (783 lbs.)
668 L (23.5 cu.ft.)
Stainless steel, Adjustable,
3 shelves, W848 x D533mm,
Max. load: 50kg (110 lbs.)/shelf
17mm diameter, 3 locations
(back, bottom left/right corner)
Hermetic type, Output: 1100 W
(high stage), 1100 W (low stage)
HFC refrigerants
High/low temperature,
Power failure, Filter check,
Battery check
Allowable contact capacity:
DC 30V, 2A
1 set of keys, 1 scraper
220V / 60 Hz
Built-in voltage booster
*Appearance and specifications are subject to change without notice.
ISO 9001 & 14001 certified
Distributed by:
Performance Data
Options
Circular recorder
MTR-C954
Strip Chart Recorder
MTR-85H
Recorder mounting
bracket
MDF-S3085
(for MTR-85H)
LCO2 backup system
CVK-UB2 (I)
LN2 backup system
CVK-UBN2
40
20
0
–20
–40
–60
–80
–100
0 2 4 6 8 10
50
12
352520
14
1510
Temperature (°C)
40
20
0
–20
–40
–60
–80
–100
0 2 4 6 8 10 12 14
Temperature (°C)
Time (hour)
Time (hour)
Pull-up characteristics
during power failure
30
12
8
6
4
2
10
0
Time (hour)
Ambient temperature (°C)
CO
2
backup system operation
characteristics
Central air temp. in cabinet
Ambient temp. 30°C, No-load
Ambient temp. and maintenance
of –70°C by injection of LCO
2
(30kg)
Inventory rack
SUR-624-I
Rack Capacity: 24
SANYO Electric Co., Ltd.
SANYO Sales & Marketing Corporation
5-15, Hiyoshi-cho 2-chome, Moriguchi City, Osaka, 570-8634, Japan
Telephone: +81-6-6992-0322 Fax: +81-6-6992-9291
URL http://www.sanyo.co.jp/cmg/biomedical
©2006 SANYO Printed in Japan 2006.3. MA. SHR101
18
(0.7)
48
(1.9)
1130 (44.5)
870 (34.3)
1196 (47.1)
1540 (60.6)450 (17.7)
1990 (78.3)
605.5 (23.8)
105
(4.1)
1280 (50.4)
130 (5.1)130 (5.1)
140
(5.5)
795 (31.3)
867 (34.1)
11
(0.4)
1625 (64)
Features
EVOLUTIONARY DESIGN
• The newly developed refrigeration system
and the freezer structure offer a quiet
experimental environment.
• Achieves 10% higher energy conservation
than previous model.
(For model rated 220V, 50Hz)
• The newly developed back cover combined
with new aerodynamically designed and
placed components in the refrigeration
compartment provide superior air flow,
making it possible
to drastically
reduce the stress
to the freezer and
contributing to
excellent durability.
• Two independent and insulated inner doors
prevent cold air leakage.
ACCESSIBILITY
• A new beak style inner
door latch tightly closes
the inner door against
the freezer frame. It
also helps make
opening and closing the door smoother.
• The condenser filter is situated at the
bottom right side of the front panel to make
filter removal and cleaning easier.
Back
cover
Compressors
Front
Rear
Air
SAFETY
• Alarm lamp and buzzer offer secure warning
of power failure or abnormal temperature
increase.
• High and low temperature warning provides
an audible and visual alarm when the
temperature deviates more than 5°C to
20°C (adjustable) from the set point.
• Alarm ring back function ensures that buzzer
will resume operation, should alarm
conditions continue after it is silenced.
• Microprocessor-controlled filter-clog check
function protects the refrigeration circuit.
• New rugged, one-handed
outer door latch has a hole to
allow a padlock to securely
protect valuable samples.
• Control panel with
digital display for easy
operation.
Pull-down characteristics
*Cooling performance is indicated by the temperature reached at
the center of the freezer (at ambient temperature of 30C with no
load). In order to use the freezer at a stable temperature for a long
time, it is recommended that the temperature be set to at least 5C
higher than the indicated lowest temperature.
In addition, depending on the usage conditions, it may not be
possible to reach the indicated lowest temperature.
MDF-U7386S
Dimensions
[Unit : mm (inch)]
Specifications
Model
Temperature range
Maximum cooling
performance
Exterior dimensions
(W x D x H)
Interior dimensions
(W x D x H)
Net weight
Effective capacity
Shelf
Access port
Compressor
Refrigerant
Alarm
Remote alarm contact
Accessories
Power Requirement
MDF-U7386SC
–50°C to –86°C (1°C increment)
–86°C
(Ambient temp. 30°C)
1130 x 875 x 1990mm
(44.5” x 34.4” x 78.3”)
870 x 600 x 1280mm
(34.2” x 23.6” x 50.4”)
Approx. 355kg (783 lbs.)
668 L (23.5 cu.ft.)
Stainless steel, Adjustable,
3 shelves, W848 x D533mm,
Max. load: 50kg (110 lbs.)/shelf
17mm diameter, 3 locations
(back, bottom left/right corner)
Hermetic type, Output: 1100 W
(high stage), 1100 W (low stage)
HFC refrigerants
High/low temperature,
Power failure, Filter check,
Battery check
Allowable contact capacity:
DC 30V, 2A
1 set of keys, 1 scraper
220V / 60 Hz
Built-in voltage booster
*Appearance and specifications are subject to change without notice.
ISO 9001 & 14001 certified
Distributed by:
Performance Data
Options
Circular recorder
MTR-C954
Strip Chart Recorder
MTR-85H
Recorder mounting
bracket
MDF-S3085
(for MTR-85H)
LCO2 backup system
CVK-UB2 (I)
LN2 backup system
CVK-UBN2
40
20
0
–20
–40
–60
–80
–100
0 2 4 6 8 10
50
12
352520
14
1510
Temperature (°C)
40
20
0
–20
–40
–60
–80
–100
0 2 4 6 8 10 12 14
Temperature (°C)
Time (hour)
Time (hour)
Pull-up characteristics
during power failure
30
12
8
6
4
2
10
0
Time (hour)
Ambient temperature (°C)
CO
2
backup system operation
characteristics
Central air temp. in cabinet
Ambient temp. 30°C, No-load
Ambient temp. and maintenance
of –70°C by injection of LCO2 (30kg)
Inventory rack
SUR-624-I
Rack Capacity: 24
SANYO Electric Co., Ltd.
SANYO Sales & Marketing Corporation
5-15, Hiyoshi-cho 2-chome, Moriguchi City, Osaka, 570-8634, Japan
Telephone: +81-6-6992-0322 Fax: +81-6-6992-9291
URL http://www.sanyo.co.jp/cmg/biomedical
©2006 SANYO Printed in Japan 2006.3. MA. SHR101
18
(0.7)
48
(1.9)
1130 (44.5)
870 (34.3)
1196 (47.1)
1540 (60.6)450 (17.7)
1990 (78.3)
605.5 (23.8)
105
(4.1)
1280 (50.4)
130 (5.1)130 (5.1)
140
(5.5)
795 (31.3)
867 (34.1)
11
(0.4)
1625 (64)
Features
EVOLUTIONARY DESIGN
• The newly developed refrigeration system
and the freezer structure offer a quiet
experimental environment.
• Achieves 10% higher energy conservation
than previous model.
(For model rated 220V, 50Hz)
• The newly developed back cover combined
with new aerodynamically designed and
placed components in the refrigeration
compartment provide superior air flow,
making it possible
to drastically
reduce the stress
to the freezer and
contributing to
excellent durability.
• Two independent and insulated inner doors
prevent cold air leakage.
ACCESSIBILITY
• A new beak style inner
door latch tightly closes
the inner door against
the freezer frame. It
also helps make
opening and closing the door smoother.
• The condenser filter is situated at the
bottom right side of the front panel to make
filter removal and cleaning easier.
Back
cover
Compressors
Front
Rear
Air
SAFETY
• Alarm lamp and buzzer offer secure warning
of power failure or abnormal temperature
increase.
• High and low temperature warning provides
an audible and visual alarm when the
temperature deviates more than 5°C to
20°C (adjustable) from the set point.
• Alarm ring back function ensures that buzzer
will resume operation, should alarm
conditions continue after it is silenced.
• Microprocessor-controlled filter-clog check
function protects the refrigeration circuit.
• New rugged, one-handed
outer door latch has a hole to
allow a padlock to securely
protect valuable samples.
• Control panel with
digital display for easy
operation.
Pull-down characteristics
*Cooling performance is indicated by the temperature reached at
the center of the freezer (at ambient temperature of 30C with no
load). In order to use the freezer at a stable temperature for a long
time, it is recommended that the temperature be set to at least 5C
higher than the indicated lowest temperature.
In addition, depending on the usage conditions, it may not be
possible to reach the indicated lowest temperature.
MDF-U7386S
Dimensions [Unit : mm (inch)]
Specifications
Model
Temperature range
Maximum cooling
performance
Exterior dimensions
(W x D x H)
Interior dimensions
(W x D x H)
Net weight
Effective capacity
Shelf
Access port
Compressor
Refrigerant
Alarm
Remote alarm contact
Accessories
Power Requirement
MDF-U7386SC
–50°C to –86°C (1°C increment)
–86°C
(Ambient temp. 30°C)
1130 x 875 x 1990mm
(44.5” x 34.4” x 78.3”)
870 x 600 x 1280mm
(34.2” x 23.6” x 50.4”)
Approx. 355kg (783 lbs.)
668 L (23.5 cu.ft.)
Stainless steel, Adjustable,
3 shelves, W848 x D533mm,
Max. load: 50kg (110 lbs.)/shelf
17mm diameter, 3 locations
(back, bottom left/right corner)
Hermetic type, Output: 1100 W
(high stage), 1100 W (low stage)
HFC refrigerants
High/low temperature,
Power failure, Filter check,
Battery check
Allowable contact capacity:
DC 30V, 2A
1 set of keys, 1 scraper
220V / 60 Hz
Built-in voltage booster
*Appearance and specifications are subject to change without notice.
ISO 9001 & 14001 certified
Distributed by:
Performance Data
Options
Circular recorder
MTR-C954
Strip Chart Recorder
MTR-85H
Recorder mounting
bracket
MDF-S3085
(for MTR-85H)
LCO2 backup system
CVK-UB2 (I)
LN2 backup system
CVK-UBN2
40
20
0
–20
–40
–60
–80
–100
0 2 4 6 8 10
50
12
352520
14
1510
Temperature (°C)
40
20
0
–20
–40
–60
–80
–100
0 2 4 6 8 10 12 14
Temperature (°C)
Time (hour)
Time (hour)
Pull-up characteristics
during power failure
30
12
8
6
4
2
10
0
Time (hour)
Ambient temperature (°C)
CO2 backup system operation
characteristics
Central air temp. in cabinet
Ambient temp. 30°C, No-load
Ambient temp. and maintenance
of –70°C by injection of LCO2 (30kg)
Inventory rack
SUR-624-I
Rack Capacity: 24
SANYO Electric Co., Ltd.
SANYO Sales & Marketing Corporation
5-15, Hiyoshi-cho 2-chome, Moriguchi City, Osaka, 570-8634, Japan
Telephone: +81-6-6992-0322 Fax: +81-6-6992-9291
URL http://www.sanyo.co.jp/cmg/biomedical
©2006 SANYO Printed in Japan 2006.3. MA. SHR101
18
(0.7)
48
(1.9)
1130 (44.5)
870 (34.3)
1196 (47.1)
1540 (60.6)450 (17.7)
1990 (78.3)
605.5 (23.8)
105
(4.1)
1280 (50.4)
130 (5.1)130 (5.1)
140
(5.5)
795 (31.3)
867 (34.1)
11
(0.4)
1625 (64)
Features
EVOLUTIONARY DESIGN
• The newly developed refrigeration system
and the freezer structure offer a quiet
experimental environment.
• Achieves 10% higher energy conservation
than previous model. (For model rated 220V, 50Hz)
• The newly developed back cover combined
with new aerodynamically designed and
placed components in the refrigeration
compartment provide superior air flow,
making it possible
to drastically
reduce the stress
to the freezer and
contributing to
excellent durability.
• Two independent and insulated inner doors
prevent cold air leakage.
ACCESSIBILITY
• A new beak style inner
door latch tightly closes
the inner door against
the freezer frame. It
also helps make
opening and closing the door smoother.
• The condenser filter is situated at the
bottom right side of the front panel to make
filter removal and cleaning easier.
Back
cover
Compressors
Front
Rear
Air
SAFETY
• Alarm lamp and buzzer offer secure warning
of power failure or abnormal temperature
increase.
• High and low temperature warning provides
an audible and visual alarm when the
temperature deviates more than 5°C to
20°C (adjustable) from the set point.
• Alarm ring back function ensures that buzzer
will resume operation, should alarm
conditions continue after it is silenced.
• Microprocessor-controlled filter-clog check
function protects the refrigeration circuit.
• New rugged, one-handed
outer door latch has a hole to
allow a padlock to securely
protect valuable samples.
• Control panel with
digital display for easy
operation.
Pull-down characteristics
*Cooling performance is indicated by the temperature reached at
the center of the freezer (at ambient temperature of 30C with no
load). In order to use the freezer at a stable temperature for a long
time, it is recommended that the temperature be set to at least 5C
higher than the indicated lowest temperature.
In addition, depending on the usage conditions, it may not be
possible to reach the indicated lowest temperature.
DMREV606.1