Sanyo MDF-U7486S Freezer User Manual


 
- 55 -
13
START-UP OF UNIT
Follow the procedures for the initial and consequent operations of the unit.
1. Turn off the switch of the back-up system (optional component) if it is installed.
2. Connect the power supply cord to the dedicated outlet having appropriate rating with the chamber
empty, and turn on the power switch on the freezer.
3. Turn on the battery switch.
4. Set the desired chamber temperature. See page 14 for the temperature setting.
5. Check that the chamber temperature reaches the desired temperature.
6. Turn on the switch of back-up system (optional component) if it is installed.
7. Make sure that the alarm lamp blinks and the buzzer sounds by pressing the alarm test key (ALARM
TEST). The remote alarm is also operated. E09 is displayed on the digital temperature indicator if the
battery switch is OFF. Make sure to turn on the battery switch.
8. After confirming the above, you can put articles into the freezer chamber in a small batch to prevent the
temperature rise.
9. An alarm function is to work normally, or push and confirm an alarm test key (the condition that a
battery switch is on). It is normal if alarm lamp blinking, remote alarm operation and a buzzer ring.
Operation after power failure
The set value is memorized by nonvolatile memory. Accordingly, the freezer resumes the operation with
setting before power failure.
When the freezer is recovered from power failure with the chamber temperature higher than the preset
temperature, then the high temperature alarm is activated and the buzzer sounds and the remote alarm is
also activated. Please push the buzzer stop key (BUZZER) to silence buzzer and take appropriate
actions if needed.