1701 SOUTH SUTRO TERRACE
CARSON CITY, NV 89706-0364
TELEPHONE (800) 993-9309
(775) 883-2500
FAX: 775-883-6388
See us on the WEB at http://www.universalanalyzers.com
e-mail address: sales@universalanalyzers.com
510 Text Rev -
UNIVERSAL ANALYZERS MODEL 510S SAMPLE COOLER
SPECIFICATIONS
SAMPLE FLOW RATE: 0 to 2 L/M (at STP)
MAXIMUM INLET TEMPERATURE: 500
o
F. (260
o
C.)
MAXIMUM INLET GAS DEW POINT: 180
o
F. (82
o
C.)
MAXIMUM INLET WATER CONCENTRATION: 50%*
MINIMUM AMBIENT TEMPERATURE: 32
o
F. (0
o
C.)
MAXIMUM AMBIENT TEMPERATURE: 105
o
F. (41
o
C.)*
MAXIMUM COOLING POWER: 50 BTU'S/Hr. (47 kJ/Hr.)
OUTLET SAMPLE DEW POINT: 41
o
F. (5
o
C.), adjustable
MAXIMUM INPUT POWER: 150 WATTS
VOLTAGE: 90 - 125 VAC, 50/60 Hz
or
180 - 250 VAC, 50/60 Hz
ELECTRICAL CLASSIFICATION: NEMA 1
DIMENSIONS: 9" x 10" x 6", HWD
WEIGHT: 11 LB's., (5 KG)
*at reduced flow rates. See capacity chart.