SPECIFICATIONS
Power SupplySpecifications
Power supply Operating Condition Units
Voltage Range 100‑240 VoltsAC
Frequency 50/60 Hertz
Current 0.6 Amps
Environmental OperatingConditions
Environmental Condition Operating Condition Units
Altitude 3000(max) Meters
Operating Temperature 16– 29 ºC
Storage Temperature ‑10to 50 ºC
Operating Relative Humidity 20‑80(non‑condensing) %
Voltage Range 12 VoltsDC
Frequency DC
Current 1.2 Amps
Installation/Over Voltage CategoryII
Pollution Degree 2
INSTRUCTIONS FOR USE – InitialSet‑up
ChooseLocation
Place 3M™Attest™ Auto‑reader390G on a firm and level surface, away from sunlight
and incandescent light. To save counter space, units may be stacked twohigh.
WARNING:
•
Do not place the instrument in environment exposed to sunlight or strong
incandescentlight.
•
Do not place instrument close to any device that emits a strong
electro‑magneticfield.
•
Do not use on a vibrating surface.
Select & Install Correct PowerPlug
The 3M™Attest™ Auto‑reader390G power supply comes with multiple blade
configurations. Install the blade configuration appropriate for your country into the power
supply and plug into appropriate AC groundedoutlet.
WARNING: Use only the power supply specified for this product and certified for
the country ofuse.
Wait 30Minutes BeforeUsing
Once plugged in, allow a 30minute warm‑up period before placing 3M™Attest™ Rapid Readout Biological Indicators into the incubation wells. The C1caution code will disappear
when proper incubation temperature isreached.
SpecialConsiderations
It is recommended that the unit be left on to eliminate warm‑up periods. The unit may be turned off when it is anticipated that it will not be used for an extendedperiod.