A SERVICE OF

logo

3-15 PULSE MODULATION
TEST: VIDEO
FEEDTHROUGH
This pulse modulation test verifies that video feedthrough is within
specifications.
Test Setup
Connect the equipment, shown in Figure 3-12, as
follows:
1. Connect the 682XXB/683XXB rear panel PULSE
SYNC OUT to the Oscilloscope Trigger Input.
2. On the Oscilloscope, select 50W input impedance.
3. Connect the 682XXB/683XXB RF OUTPUT to the
Oscilloscope Vertical Input through a 450 MHz
Low Pass Filter.
682XXB/683XXB MM 3-43
PERFORMANCE PULSE MODULATION TEST:
VERIFICATION VIDEO FEEDTHROUGH
450 MHz
Low Pass Filter
(For Video Feedthrough
Measurement)
CH. 4 Trigger
Pulse
Sync
Out
682XXB / 683XXB
CH. 1 Input
Oscilloscope
Figure 3-12. Equipment Setup for Pulse Modulation Test: Video Feedthrough