Model 5328A
Maintenance
Table 5-8C. HP-IR Bus Commands
Check
Point
Test
17
LOCAL
When the 9825A CONTINUE key is pressed, verify that counter (RMT)
LOCKOUT
annunciator is on.
(LLO)
When the 9825A CONTINUE key is again pressed, verify Local Lockout
by pressing front panel RESET button and ensuring counter doesn’t go into
Local operation. (RMT annunciator Off).
DEVICE CLEAR
When the 9825A CONTINUE key is pressed, verify that counter displays
(DCL)
10.000 MHz. When the 9825A CONTINUE key is again pressed, counter
will reset to its Remote Program Initialize mode and display (O.).
SELECTED
When the 9825A CONTINUE key is pressed, verify that counter displays
DEVICE
10.000 MHz. When the 9825A CONTINUE key is again pressed, counter
CLEAR (SDC)
will reset to its Remote Program Initialize mode and display (O.).
GROUP
When the 9825A CONTINUE key is pressed, verify that counter is in Hold
EXECUTE
(Gate Light off). When the 9825A CONTINUE key is pressed, Gate Light
TRIGGER (GET)
should flash twice and counter should display 10.000 MHz.
SERIAL POLL
When the 9825A CONTINUE key is pressed, counter should display
(SPE/SPD)
(0.0000000 MHz) and GATE Light should go off. Calculator should print
(STATUS BYTE = 64.00).
GO TO LOCAL
When the 9825A CONTINUE key is pressed, verify that counter is in Local
(GTL)
(RMT annunciator off). When 9825A CONTINUE key is again pressed,
counter will go into remote.
END OF TEST
Figure 5-9. Triangular Pulse Oserved in Steps 5 and 7, Table 5-8B
5-38