![](https://pdfstore-manualsonline.prod.a.ki/pdfasset/1/f7/1f780004-10e9-46e2-b035-e6989e4ee288/1f780004-10e9-46e2-b035-e6989e4ee288-bg1a.png)
SECTION 4—MAINTENANCE
Invacare
®
Pediatric Bear 26 Part No 1160837
Nebulizer Change
Thenebulizerisdesignedforsinglepatientuse.Thenebulizer
MUSTbereplacedafteralongperiodofinactivity,orwhenit
isobstructedbydrymedication,dust,etc.Reusable
nebulizersshouldlastfromsixmonthstooneyearwith
generaluse.Refertothenebulizersowner’smanual.
Disposablenebulizers shouldlast30days.Fordisposable
nebulizers,itisrecommendedthatanextranebulizerbekept
atalltimes.Refertothedisposablenebulizer’sowner’s
manual.
Filter Change
1. ThefilterMUSTbereplacedafterapproximately30hours
ofuseorwhenitturnsgrey.
2. Openthefilterholdertoreplacethefilterwithanewone.
CAUTION
DONOTusecottonoranyothermaterial.
DONOTwashorcleanthefilter.
UseonlyInvacarefilters.
DONOToperateunitwithoutanairfilter.