A SERVICE OF

logo

20 ST60 Depth Instrument Owners Handbook
3. Apply the flush mount template (supplied at the rear of this handbook) to the
selected location and mark out the aperture into which the assembled instru-
ment and bezel will sit.
4. Cut out the aperture (3) for the assembled instrument and bezel and remove
the template.
5. Peel off the protective sheet from the self-adhesive gasket (4) then stick the
gasket into position on the rear of the bezel.
6. Screw the two fixing studs (1) into the threaded sockets on the rear of the
instrument.
7. Mount the assembled instrument, studs, bezel and gasket into the panel.
8. Locate the flush mount bracket (6) onto the fixing studs and secure the assem-
bly to the panel with the thumb-nuts (5).
Bracket mounting
A Control Unit Mounting Bracket (Part No. E25009) enables you to mount your
ST60 instrument in locations where other forms of mounting are impractical.
Although this provides a useful alternative method for securing your instrument,
it is only suitable for use in positions where the instrument will not be exposed to
water.
To bracket mount your ST60 instrument, do so in accordance with the Control Unit
Mounting Bracket Instruction Sheet.
Fitting transducer
The ST60 Depth instrument is supplied, with a through-hull depth transducer.
The depth transducer is supplied with detailed instructions for installation and
maintenance. Before attempting to install the depth transducer, read these
instructions and the
Site requirements
for transducers described in this Chapter.
Once you are satisfied you can meet all the installation requirements, install the
transducer in accordance with the accompanying installation instructions.
Running transducer cable
Each transducer type has a 14 m (45 ft) cable fitted with spade terminals for
connection to the ST60 Depth instrument. The manner in which you run the cable
will depend on the locations of the transducer and instrument.
Observing the following guidelines, run the transducer cable to the ST60 Depth
instrument:
If the cable has to be fed through the deck, always use a proprietary deck
gland.
81038_4.book Page 20 Wednesday, March 24, 2004 3:17 PM