A SERVICE OF

logo

14
Connection on PCB
Following is the explanation of connections on Main PCB.
Connector Connects to Usage Voltage
CN1 Switching power supply To supply the power to PCB #1:12V
CN2
MTR-480 Option
To connect with interface
CN3
Remote alarm terminal
#1: COM.
#2: N.O.
#3: N.C.
Output for remote alarm contact
CN4
#1-#2: Temp. control relay
#5-#6: Fan relay
To control chamber temperature
To conduct current to cap. tube
heater
#1-#2:12V
#5-#6:12V
CN5
#1, #7: Buzzer
#2-#6: Control PCB (CN51)
To control with each switches
CN6
Display PCB CN52
To connect with each LED
CN7 #7-#8: Comp. sensor
To control condensing fan motor
by the temperature in comp.
sensor
CN8 #1-#2: Battery
To supply the power to Main PCB
during power failure
#2:6V
CN9 Unused
CN11 #1-#3: Temp. sensor To detect chamber temperature