AC suppression
response time
Instantaneous
Dataline suppression 2-Line Tel/DSL
EMI / RFI AC noise
suppression
Yes
PHYSICAL
Shipping weight (lbs) 27.4
Shipping weight (kg) 12.3
Unit weight (lbs) 25.4
Unit weight (kg) 11.4
Unit Dimensions
(HWD/in)
10.6 x 5.2 x 7.4
Unit Dimensions
(HWD/cm)
27 x 13.2 x 18.8
Material of
construction
PVC
Form factors supported Tower
Cooling method Convection
Battery Access Battery access door allows hot-swap battery replacement without powering connected equipment off
ENVIRONMENTAL
Relative Humidity 0 to 95%, non-condensing
COMMUNICATIONS
Network monitoring
port
USB communications port
Software and cabling
included
Ships with USB cable, PowerAlert software available via free download from www.tripplite.com
LINE / BATTERY TRANSFER
Transfer time from line
power to battery mode
2-4 milliseconds
Low voltage transfer to
battery power
159V (resets to line power as voltage increases to 167V)
High voltage transfer
to battery power
266V (resets to line power as voltage decreases to 259V)
SPECIAL FEATURES
Cold Start Yes, inverter can be "cold started" to enable temporary AC output during a power failure
Appearance Attractive gray tower UPS
CERTIFICATIONS
Certifications Tested to CE/ IEC 60601-1 (Series AGOM6832), Class B (emissions)
WARRANTY
Product Warranty
Period
2 year product warranty
Optional Coverage
More information, including related products, owner's manuals, and additional technical specifications, can be found online at
www.tripplite.com/en/products/model.cfm?txtModelID=4027.