Standard & Optional Accessories
Standard
Test leads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ATL55
Instruction manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .DL250-MAN
Temperature probe connection adapter . . . . . . . . . . . . . . . .ATT70
Thermocouple . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ATT29
Optional
Flame safeguard test kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ATLFSG
Microwave diode booster test lead . . . . . . . . . . . . . . . . . . . .ATL60
Plug-in line splitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ALS1
Hard carrying case (service kits) . . . . . . . . . . . . . . . . . . . . . .AC506
Digital tachometer adapter . . . . . . . . . . . . . . . . . . . . . . . . . .DPM2K
The specifications listed in the following table are based on an instru-
ment operating in an ambient temperature of between 65 and 90˚F, at a
relative humidity of less than 80%, and within one year of calibration.
The “accuracy” column indicates the percent of possible error based on
the reading displayed, plus an additional allowance for error based on
the value of the least significant digits of the selected range. This is stat-
ed as (±X% + X dgts).
AC readings displayed on this meter are average responding, True-RMS
indicating. They are based on a true sinusoidal waveform.
DL250-MAN P. 9
Maximum voltage between any 600 V DC or AC RMS
terminal and earth ground
Digital display 4000 Count - Updates 4 times
per second
Storage temperature -4˚ to 140˚F (-20˚ to 60˚)
Operating temperature 32˚ to 113˚F (0˚ to 45˚C)
Altitude = or < 6560’ (2000 M)
Relative humidity 0% to 80% at 32˚ to 95˚F (0˚ to 35˚C)
0% to 70% at 95˚ to 113˚F (35˚ to 45˚C)
B a t t e ry ty p e 9V, NEDA, 1604 or 6LR 61
B a t t e ry life 80 hours (alkaline)
M aximum conductor size 1. 25” (32 mm) up to 750 MCM
M aximum jaw opening 1. 25” (32 mm)
Size (H x W x L, in mm) 235 x 86.5 x 45
Weight (approximate) .95 lbs (430 g)
V i b ration and shock Designed to MIL -T 28800 for a
class II instrument
Case protection IEC 529, IP 30
S a f e ty s tandards Designed to UL 3III (overvoltage
ca t e g o ry III) CSA C 22.2 no 1010 - 1,
and both IEC 1010 - 2 - 0 32, and
the EMC directive
Function
AC Amps
( 5 0 Hz to 400 Hz)
D C
Low Amps
DC Vo l t s
AC Vo l t s
( 5 0 Hz to 400 Hz)
O h m s
C a p a c i t a n c e
Te m p e r a t u r e
Range
40 A
400 A
40 µA
400 µA
4 V
40 V
400 V
600 V
4 V
40 V
400 V
600 V
400 Ω
4K Ω
40K Ω
400K Ω
4 Ω
40 Ω
1 µF
10 µF
100 µF
1000 µF
10000 µF
-40˚ to 15˚F
15˚ to 750˚F
Resolution
0.01 A
0.1 A
0.01 µA
0.1 µA
1 mV
10 mV
100 mV
1 V
1 mV
10 mV
100 mV
1 V
100 Ω
1 Ω
10 Ω
100 Ω
1 Ω
10 Ω
0.001 µF
0.01 µF
0.1 µF
1 µF
1 µF
0.1˚F
0.1˚F
Accuracy
3.0% ±20 dgts
2.0% ±5 dgts
1.0% ±5 dgts
1.5% ±2 dgts
0.9% ±2 dgts
0.9% ±3 dgts
0.9% ±3 dgts
1.2% ±3 dgts
1.5% ±5 dgts
1.7% ±5 dgts
2.5% ±15 dgts
15% ±100 dgts
5% ±5˚F
1% ±3˚F
Overload
400 A
continuous
400 µA/250 V
600 V RMS
600 V
600 V
30 V AC or
60 V DC
C o n t i n u i ty Open circuit test voltage <2.7 V 600 V
Threshold: Approx: < 50 Ω
Diode check Open circuit test voltage <3.3 V 600 V
M ax test current 3.0 mA