8)ONLINE ILLEGAL SAMPLE KIND No. [ aa ]
<Operation>
(1).Disregard the test requisition information text for the sample generating the alarm.
(2).Continue to receive the test requisition information text for the other samples within the run.
<Operation>
(1).When receiving the test requisition information text using online, sample kind No. of the sample generating the alarm
does not meet the specification.
(2).The display of [ ] and its meaning are as follows.
aa : Sample type
----------------------------------------------------------------
__ : Normal sample
_E : Emergency sample
_P : Stat sample
U_ : Urine normal sample
UE : Urine emergency sample
UP : Urine stat sample
X_ : Other normal sample
XE : Other emergency sample
XP : Other stat sample
----------------------------------------------------------------
9)ONLINE ILLEGAL GENDER TEXT [ a ]
<Operation>
(1).Disregard the test requisition information text for the sample generating the alarm.
(2).Continue to receive the test requisition information text for other samples within the run.
<Description>
(1).When receiving the requisition information text using online, the sex of this text is not within the specification.
(2).The display of [ ] and its meaning are as follows.
a : Received sex
10)ONLINE ILLEGAL AGE/MONTH [ aaa bb ]
<Operation>
(1).Disregard the test requisition information text for the sample generating the alarm.
(2).Continue to receive the test requisition information text for other samples within the run.
<Description>
(1).When receiving the requisition information text using online, age or month age (age for baby before one year old) of this
text is out of the specification.
(2).The display of [ ] and its meaning are as follows.
aaa : Received age
bb : Received month age