2Wire AU2700 Coffeemaker User Manual


 
- 65 -
16)ONLINE RERUN ITEM ERROR [ aaaaaaaa ]
<Operation>
(1).Disregard the repeat run test requisition information for the sample generating the alarm.
(2).Continue to receive the test requisition information text for the other samples within the run.
<Description>
(1).When receiving the test requisition information text for repeat run using online, repeat run test requisition cannot be
stored with for the following reasons.
1).Original sample that is designated in the repeat run text is not stored in the normal run/sample test requisition
information.
2).Sample No. is not set to the original sample that is designated in the repeat run text.
3).Original sample that is designated in the repeat run text has already been stored as a repeat run sample
for another sample.
(2).The display of [ ] and its meaning are as follows.
----------------------------------------------------------------
aaaaaa : sample type and original sample No.
__nnnn : Normal sample
__Ennn : Emergency sample
__Pnnn : Stat sample
_Unnnn : Urine normal sample
_UEnnn : Urine emergency sample
_UPnnn : Urine stat sample
_Xnnnn : Other normal sample
_XEnnn : Other emergency sample
_XPnnn : Other Stat sample
----------------------------------------------------------------