PERFORMANCE VERIFICAITONS
Perform the following analysis, if the meter conforms to the limits listed in Table 1 the meter is functioning
correctly. If the meter does not conform to any of the listed limits the calibration procedure must be
performed.
Performance Verification Preparation
1. Turn on the Calibrator, allow calibrator to warm up. Temperature Stabilization
should be reached after 30 minutes.
2. Remove battery cover and using a calibrated meter to ensure the battery measures a minimum of 7.5 V
DC. If the battery measures under 7.5V DC, replace the battery before beginning the performance test.
3. Input the values listed in Table 1 for the Standard 61-362
Table 2 Performance Verification 61-362
Function Setting /Range Input Low Limit High Limit
ACV 3.2V 3.000V @ 50Hz 2.936 3.064
ACV 3.2V 3.000V @ 500Hz 2.936 3.064
ACV 32 30.00 @ 50Hz 29.36 30.64
ACV 32 30.00 @ 500Hz 29.36 30.64
ACV 320 300V AC @ 50Hz 293.6 306.4
ACV 320 300V AC @ 500Hz 293.6 306.4
ACV 600 500V AC @ 50Hz 486 514
ACV 600 500V AC@ 60Hz 486 514
DCV 320mV 300mV DC 296.3 303.7
DCV 3.2V 3.000V DC 2.975 3.025
DCV 32V 30.00V DC 29.63 30.37
DCV 320 300.0V DC 296.3 303.7
DCV 600 500V DC 493 507
A DC 320µA 300µA 293.9 306.1
A DC 3200 µA 3000µA 2939 3061
A DC 32mA 30mA 29.39 30.61
A DC 320mA 300mA 293.9 306.1
A DC 10 9.00 DCA 8.70 9.30
A AC 320µA 300µA @ 50Hz 292.1 307.9
A AC 320µA 300µA @ 500Hz 292.1 307.9
A AC 3200µA 3000µA @ 50Hz 2921 3079
A AC 320oµA 3000µA @ 500Hz 2921 3079
A AC 32mA 30mA @ 50Hz 29.21 30.79
A AC 32mA 30mA @ 500Hz 29.21 30.79
A AC 320mA 300mA @ 50Hz 292.1 307.9
A AC 320mA 300mA @500Hz 292.1 307.9
A AC 10A 9A @ 50Hz 8.64 9.36
A AC 10A 9A @ 500Hz 8.64 9.36
320
300.0 297.7 306.3
2k
3.000k 2.950 3.048
32k
30.00k 29.52 30.48
320k
300.0k 295.2 304.8
2M
3.000M 2.922 3.078
32M
30.00M 28.45 31.55
Diode Test 500mV .448 .552
Continuity Test
120
Beep off
Continuity Test
80
Beep on
Form number 61362 rev 1 May 2002