HART Mode Users Guide
6
Supported vs. Generic Transmitters
The calibrator communicates with virtually all HART
transmitters. In addition, the calibrator is programmed to
use device-dependent commands for a selection of HART
transmitters and their software versions. These are
“supported transmitters.” All other transmitters are
“generic.”
Table 1 shows the operations that are available for
supported vs. generic transmitters in single point and
multidrop configurations.
Note
Sensor trim is provided for supported
transmitters, with a few exceptions, as identified
in the list of supported transmitters that you can
view on the display.
To display a list of supported transmitters and software
versions:
1. If the calibrator is connected to a HART transmitter
with the Active Device screen showing, press the
Abort softkey, followed by More Choices.
2.
If the calibrator is not connected to a HART
transmitter, press the r key followed by
More
Choices
.
3.
Press the
Device Revs softkey. The Browser screen
appears.
pe07s.bmp
4. Press u or d to highlight the desired manufacturer,
and press e. A list of model numbers appears.
5. Press u or d to highlight the model number, and
press e. A list of software versions appears.