Haier DW-86L626 Freezer User Manual


 
20
1.Temperature recorder
The ULT Freezer is equipped with a 6-inch temperature recorder for the chamber
temperature. When the unit is powered up, the recorder starts working.
For proper operation of the temperature recorder, please follow the procedures below.
1.Open the recorder door to access the recorder.
2.Plug in the 9 volt battery located on the upper right corner of the recorder. This
battery is a standby power.
3.Install a new chart by following the diagram below.
Optional Accessories
2.Power supply
The temperature recorder requires an AC power supply. If AC power supply is
interrupted, the LED of the recorder blinks and the warning indicator turns on. The
temperature recorder can also be operated with DC from spare source battery which
lasts about 30 hours.(Caution: The battery should remain fully charged. If the unit is not
in operation, the battery should be switched off. Ensure that the battery is in good
status as a spare power source.) If power fails, the LED of the temperature recorder
blinks until AC power is supplied or spare power source is connected. The LED
indicator will blink when the power in the battery is low.
Replacement of Recorder Battery
When the green LED on the recorder flashes, there are two ways to inspect the unit.
1)
Supplementary power supply (9 volt DC)
If the LED light is off and the recorder stops working, it would suggest that there might
be issues associated with the main power supply. Please inspect whether there is a
loss of power, a short circuit, or a loose connection. If there is no issue with the main
power supply and the recorder starts working normally, please install the backup
battery to the recorder. At this time, if the LED stops flashing, the supplementary
power supply is normal. If the LED continues to flashes, it means that the power supply
is under capacity requirement. It is necessary to replace the battery. The LED green
light is off when there is sufficient power to the recorder.
9V battery
Button
Hand
Dateline
Screw & fixing
Recorder
(Figure 5)
The Temperature Recorder
Button
5
If there is leakage of petroleum gas or another flammable gas, close the gas supply
valve and open doors and windows to ventilate the air. Do not plug or unplug your
freezer unit.
While the equipment is repaired, take precautions not to inhale any chemicals or
aerosols existing near the unit. They might be harmful to your health.
Never directly pour water into the unit. The water may cause electric shock or
short circuit.
Customers must not dismantle, repair or modify the equipment. Such improper
operations may result in fire or personal injury.
Should there be a malfunction in the equipment, power off the unit and unplug the
power cord from power supply. Continuous operation with an abnormal condition
may result in electric shock and fire.
In the event there is a power outage and recovery, be sure to wait for 5 minutes
before the unit power is turned back on. This is to reduce the chance of damage to
the compressors and refrigeration system. During a repair operation, power must
be disconnected. The power cord should always be protected to avoid damages.
When unplugging the power cord from the socket, use a firm grip on the plug itself
for pulling. Do not pull the power cord as this action may strip the wires off the plug
causing electric shock, electrical short circuit and fire.
Maximum loading on the each shelf should be not more than 50kg and total loading
for whole unit should be not more than 200 kg.
.
Heavier loads may cause damage to
the shelving system
Proper Usage