Haier HYC-610 Refrigerator User Manual


 
Descriptions of Refrigerator Parts and
Their Functions
1)Control Panel
2)Light Power Switch
3)Condensation Heater Control
4)Battery Power Switch
5)Door Lock
6)Door Handle
7)Casters
8)Automatic Temperature Recorder (optional)
9)Leveling Leg
10)Shelves
11)Interior Lights
12)Door Ajar Alarm
The control panel on the Haier blood bank shows the actual storage temperature inside the refrigerator.
You can use the push buttons on the panel to adjust the temperature set point and test the alarm
functions.
You can use the switch to turn on and off the interior lights of the refrigerator.
You can use this switch (located inside the back of the refrigerator behind a cover) to turn on the
heater that reduces the condensation build up. “1” means power on and “0” means power off.
The switch controls the rechargeable battery that powers the alarm system in case of the loss of main
power. “1” is on and “0” is off.
You can use this feature to lock the refrigerator.
For ease of opening and closing the door.
The casters underneath the unit provide the convenience of moving the unit around.
The refrigerator is equipped with an automatic temperature recorder. Its paper chart runs for a full turn
in 7 days, so the storage temperature can be permanently recorded. The chart paper must be
changed once every 7 days.
The unit must be level to operate. You can adjust the screws on the leveling leg underneath the
refrigerator to adjust the height.
The drawers and shelves are designed to conveniently store and unload the blood samples.
When the door is open for 10 minutes, the door ajar alarm will sound. You should reduce the open
door time to stabilize the refrigerator temperature.
15 Watts light hulb.
Maintenance
Warning
Before any inspection or maintenance work is performed, the refrigerator's power plug should be
disconnected from the power supply socket. This is to prevent any potential electrical shock or injury.
During the maintenance work, do not breathe the dust and aerosols near the unit; they might be
harmful to your health.
Cleaning the refrigerator
Changing the light bulb starter
Purchasing chart paper for temperature chart recorder
Purchasing backup battery for temperature recorder
Changing fuse breaker
1.The refrigerator should be cleaned once a month. Regular cleaning can keep the unit looking new.
2.Use a dry, soft cloth to clean the dust off the interior and exterior. If necessary, use a soft cloth with a
solution of water and mild detergent to wipe the unit.
3.After the cleaning, use a dry cloth to wipe off any solution residue off the surfaces.
4.Do not pour water directly into the refrigerator. By doing so, the water can damage the insulation
materials and cause problems.
5.Parts in the refrigeration system for this refrigerator are completely sealed. They do not require any
lubricants.
6.The air filter for the condenser should be cleaned at least once within a 90 day period. The procedure
is as follows.
a.Remove the front louver panel for the refrigerator.
b.Remove the air filter located in front of the condenser; wash it in water.
c.After cleaning the filter, install it in front of the condenser and the louver panel.
The florescent light bulb is located in the upper front area of the refrigerator. When changing the light
bulb, please perform the following steps.
1.Unplug the refrigerator from the power supply.
2.Remove the left panel and the cover for the electrical control box.
3.Remove the defective starter.
4.Install a new starter. Make sure it is securely in position.
5.Install the cover for the electrical box and the left panel.
6.Plug the power plug back into the system.
The chart recorder papers are imported chart papers designed for the recorder. The papers that are
supplied with the unit usually last about half of a year. When you are running short of the chart paper,
please contact Haier to purchase more. The lead time to deliver the charts is 15 days after receiving the
payment. Please call Haier for details.
You can purchase a high quality battery from a store for the same size and voltage (9V). You may also
call Haier to purchase a high quality replacement battery. The lead time to deliver the battery is 15 days
after receiving the payment. Please call Haier for details.
If there is a burnt out fuse on a PC board, please contact Haier's after sales support department.