46
Technical data
G 7835 CD, G 7836 CD
Washer-disinfectors G 7835 CD G 7836 CD
Front-loading unit with bottom-hinged door, excl. baskets • •
Built-under/freestanding unit with lid • –
Freestanding unit with lid –•
Freshwater system, max. temperature 93°C • •
Circulation pump [Qmax. l/min] 400 600
Controls/programmes
Freely programmable PROFITRONIC controls • •
Electric door lock ••
Peak-load negotiation ••
Serial interface for process documentation • •
Magnetic strip for automatic mobile unit recognition • •
Water supply connection
1 x cold water, 0.5–10 bar flow pressure (50–1000 kPa) • •
1 x cold water for steam condenser, 0.5–10 bar flow pressure (50–1000 kPa) • •
1 x hot water, 0.5–10 bar flow pressure (50–1000 kPa) • •
1 x demineralised water, 0.5–10 bar flow pressure (50–1000 kPa) • •
Feed pump for demineralised water (depending on model) • •
4 inlet hoses
1
/
2
" with
3
/
4
" threaded union, length approx. 1.7 m • •
Drain pump DN 22, head height 100 cm • •
Steam condenser water drain (DN 22) • •
Waterproof system (WPS) ••
Electrical connection
3 N AC 400 V 50 Hz, supply lead approx. 1.7 m, 5 x 2.5 mm
2
••
Heating [kW] 9.0 9.0
Circulation pump [kW] 0.7 1.2
Total rated load [kW] 9.7 10.2
Fuse rating [A] 3 x 16 3 x 16
Dispenser systems
1 dispenser pump for liquid acidic agents • (peristaltic pump) • (bellows-type pump)
1 dispenser pump for liquid detergent • (peristaltic pump) • (bellows-type pump)
Drawer for 2 x 5 l supply containers • –
Drawer for 4 x 5 l supply containers – •
Connections
DOS G 10 dispenser for liquid agents (surfactant) • –
DOS G 60 or DOS G 60/1 dispenser for liquid detergent • –
DOS 2 dispenser pump for liquid agents (surfactant), – •
Bellows-type pump, retrofittable by Service
DOS 4 dispenser pump for liquid agents (disinfectant), – •
Bellows-type pump, retrofittable by Service
G 7835 CD