Sanyo MDF-793 Freezer User Manual


 
14
INSTALLATION
Installation
1. Remove the packaging materials and tapes
Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If
the outside panels are dirty, clean them with a diluted neutral dishwashing detergent. (Undiluted
detergent can damage the plastic components. For the dilution, refer to the instruction of the
detergent.) After the cleaning with the diluted detergent, always wipe it off with a wet cloth. Then
wipe off the panels with a dry cloth.
2. Adjust the leveling foot
Extend the leveling feet by rotating them
counterclockwise to contact them to the floor.
Ensure the unit is level.
3. Fix the unit
Two fixtures are attached to the rear of the frame. Fix the frame to the wall with these fixtures and
rope or chain.
If holes can be opened on the wall, open a hole with a 10.5mm diameter and fix the frame by using the
chain supplied with this unit.
4. Ground (earth)
WARNING
Use a power supply outlet with ground (earth) to prevent electric shock. If the power supply outlet
is not grounded, it is necessary to install a ground by qualified engineers.
Never ground the unit through a gas pipe, water main, telephone line or lightning rod. Such
grounding may cause electric shock in the case of an incomplete circuit.