18
OPERATING INSTRUCTIONS
Setting of alarm resume time
The alarm buzzer is silenced by pressing BUZZER key on the control panel during alarm condition (Initial
setting 100: The remote alarm is not silenced).
The buzzer will be activated again after certain suspension if the alarm condition is continued. The
suspension time can be set by following the procedure shown in the Table 3 below.
The example in the table is based on the assumption that the desired duration is 30 minutes.
Note: The duration is set to 100 (no alarm resume time) at the factory.
Table 3. Setting procedure (change to 30 minutes from no alarm resume time)
Description of operation Key operated Indication after operation
1 -----
The current chamber temperature is
displayed.
2
Press the numerical value shift key
for about 5 seconds.
The first digit blinks.
The settable digit is shifted.
3
Set to F25 with the digit shift key
and numerical value shift key.
When pressed, the figure of settable
digit changes.
4 Press ENT key. ENT
The current setting is displayed
and the first digit blinks.
The settable digit is shifted.
5
Set to 130 with the digit shift key
and numerical value shift key.
When pressed, the figure of settable
digit changes.
6 Press ENT key. ENT
Alarm resume time is memorized
and the current chamber
temperature is displayed.
• The settable alarm resume time is between 10 and 60 minutes with 1 minute increment. (The setting is
010, 111, ------- 159, 160). The buzzer would not reset if the resume time is set in 000.
• The buzzer is ON when the resume time is changed in F25 while the resume time is counted (the
buzzer is silenced by BUZZER key). The count of resume time is stopped.
• The setting cannot be changed during power failure.
• The set mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation. In this case, any setting before pressing ENT key is not memorized.
• It is selectable whether the buzzer is in conjunction with the remote alarm or not when the buzzer is
silenced by BUZZER key. (The remote alarm is not in conjunction with the buzzer by factory setting.)
• The remote alarm is activated in conjunction with the buzzer if the third digit “1” is set to “0” in the step 5
above