- 47 -
9
REFRIGERATOR COMPONENTS
1. Front cover: A front cover opens when a bottom part is pulled to the front. Open this cover when you
use cleaning of the condenser filter, a remote alarm terminal. Refer to page 11.
2. Control panel: Panel opens when the right top of the lower part cover is pushed. Refer to page 10.
3. Light switch: This switch is used for turning the fluorescent lamp off and on.
4. Space for temperature recorder: An temperature recorder (optional accessory) can be mounted here.
See page 32.
5. Fluorescent lamp: 40 W white lamp. See page 24 for replacement.
6. Lock: Turn the key clockwise through 180
o
to securely lock the door.
7. Glass window: The window may have condensation in high humidity environment. Wipe off the
condensation with a soft dry cloth.
8. Handle: Always hold the handle when opening/closing the door.
9. Dew receiver: The condensation on the outer door is accumulated on this dew receiver. Wipe off the
water on the dew receiver with a cloth.
10. Leveling foot: Adjust the height of the leveling feet by turning the screw bolts until the unit is level.
11. Caster: When installing the unit, the castors can be raised from the floor by using the leveling feet
above.
12. Drawer (MPR-1411R only): Items to be stored in the chamber must be placed on the drawers.
The maximum storage weight for each drawer is 40 kg.
Shelf (MPR-1411 only): Items to be stored in the chamber must be placed on the shelves. The
maximum storage weight for each shelf is 50 kg. Do not put stored items directly onto the interior floor of
the chamber. See page 15.
13. Access port: This port allows temperature measurement cables to enter the chamber from outside.
A total of three (3) ports are provided; left side, right side and top.
14. Exhaust air vent: Do not block this vent. If this vent is blocked, temperature regulation will become
unstable. Do not place stored items in the path of the cold air.
15. Intake air vent: Do not block this vent. If this vent is blocked, temperature regulation will become
unstable. Do not insert a finger or any foreign object into this vent as there is danger from the internal fan.
16. Door switch: When a door opens, a fan for the cold circulation is stopped, and the leakage of the
cold is prevented. And, the check lamp of the door is turned on.
17. Fixture: 2 fixtures are attached on the rear side of the unit. These keep adequate space between
the unit and wall and also can be used for fixing the unit. See page 13.
18. Evaporating tray: Defrost water from the evaporator accumulates on the tray and evaporates into
the atmosphere. See page 23 for cleaning.