OPTIONAL COMPONENTS
Setting for battery before starting
Always perform the setting shown below after the installation of battery for power failure alarm, an
optional component (MPR-48B).
The chamber temperature and “E09” is displayed on the temperature display alternately if the battery
switch is turned on without following procedure.
㩵 Turn off the battery switch, perform the following setting, and turn on the battery switch again if “E09” is
displayed on the temperature display.
Description of operation Key operated Indication after operation
1-----
The current chamber temperature is
displayed.
2
Press numerical value shift key for
5 seconds.
F00 is displayed and the first digit
blinks.
3
Press numerical value shift key
twice.
The display is changed to F02 from
F00.
4 Press digit shift key once. The second digit blinks.
5
Press numerical value shift key 4
times.
The display is changed to F42 from
F02.
6 Press set key. SET
The current setting (000) is displayed
and the first digit blinks.
7
Press numerical value shift key and
scroll the figure to 1.
The display is changed to 001 from
000.
8 Press set key. SET
The setting is memorized and
the current chamber temperature is
displayed.