16
CHAMBER TEMPERATURE SETTING
Table 1 shows the basic procedure for setting the chamber temperature. Perform key operations in the
sequence indicated in the table. The example in the table is based on the assumption that the desired
temperature is 10
o
C.
Note: The unit is set at the factory with a chamber temperature of 5
o
C.
Table 1 Basic operation sequence (Example: Chamber set temperature 10
o
C)
Description of operation Key operated Indication after operation
1 Connect the power supply plug. ----
The current chamber temperature is
displayed.
2 Press set key. SET
The second digit is flashed.
By pressing key, the settable digits
moves.
3
Set to 010 with the numerical value
shift key and digit shift key.
When pressed, the figure of settable
digit increases.
4 Press set key. SET
Set temperature is memorized and the
current chamber temperature is
displayed.
Note:
• The temperature set mode returns to the temperature display mode automatically when 90 seconds has
passed without any key operation. In this case, the chamber temperature setting is not accepted.
• The available set range of chamber temperature is between 2 and 23
o
C. The partial freezing may
occurred if the set temperature is lower than 3
o
C.