Chapter 8 Specifications
Supplemental Characteristics
152
Programming Language
SCPI (Standard Commands for Programmable Instruments)
State Storage Memory
Five (5) user-configurable stored states
Recommended Calibration Interval
1 year
Output Terminal Isolation (maximum, from chassis ground)
±60 Vdc when connecting shorting conductors without insulation between the
(+) output and the (+) sense terminals and between the (-) output and the (-)
sense terminals.
±240 Vdc when connecting insulated shorting conductors between the (+)
output and the (+) sense terminals and between the (-) output and the (-) sense
terminals.
AC Input Ratings (selectable via two switches on PC board)
std 115 Vac ± 10%, 47 to 63 Hz
opt 0E3 230 Vac ± 10%, 47 to 63 Hz
opt 0E9 100 Vac ± 10%, 47 to 63 Hz
Maximum Input Power
Cooling
Fan cooled
Operating Temperature
0 to 40 °C for full rated output.
E3640A E3641A E3642A E3643A E3644A E3645A
105 VA 105 VA 205 VA 175 VA 300 VA 270 VA