A SERVICE OF

logo

Appendix Service Information
Performance Test Record for Your Power Supply
180
CC Performance Test Record
Test Description Models
Actual
Result
Specifications
Upper Limit Lower Limit
CC Programming Accuracy
@ 0 amps (I
O
)
all +0.0100 A -0.0100 A
CC Readback Accuracy @ 0 amps all I
O
+ 0.0050 A I
O
- 0.0050 A
CC Programming Accuracy
@ Full Scale (I
O
)
(E3640A) 3.01600 A 2.9840 A
(E3641A) 0.8116 A 0.7884 A
(E3642A) 5.02 A 4.98 A
(E3643A) 1.4128 A 1.3872 A
(E3644A) 8.026 7.974 A
(E3645A) 2.2144 2.1856 A
CC Readback Accuracy @ Full Scale (E3640A) I
O
+ 0.0095 A I
O
- 0.0095 A
(E3641A) I
O
+ 0.0062 A I
O
- 0.0062 A
(E3642A) I
O
+ 0.0125 A I
O
- 0.0125 A
(E3643A) I
O
+ 0.0071 A I
O
- 0.0071 A
(E3644A) I
O
+ 0.0170 A I
O
- 0.0170 A
(E3645A) I
O
+ 0.0083 A I
O
- 0.0083 A
CC Load Effect (Load Regulation) (E3640A) Maximum change: < (0.55 mA)
(E3641A) Maximum change: < (0.33 mA)
(E3642A) Maximum change: < (0.75 mA)
(E3643A) Maximum change: < (0.39 mA)
(E3644A) Maximum change: < (1.05 mA)
(E3645A) Maximum change: < (0.47 mA)
CC Source Effect (Line Regulation) (E3640A) Maximum change: < (0.55 mA)
(E3641A) Maximum change: < (0.33 mA)
(E3642A) Maximum change: < (0.75 mA)
(E3643A) Maximum change: < (0.39 mA)
(E3644A) Maximum change: < (1.05 mA)
(E3645A) Maximum change: < (0.47 mA)
CC PARD (Normal mode) all < 4 mA rms
CC PARD (Common mode) all < 1.5
mA rms