2Wire AU5400 Coffeemaker User Manual


 
- 54 -
2)ONLINE ERROR [ aa ] [ bbbbbbbbb ]
<Operation>
(1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "STOP",
the following events can occur:
1). The text following the alarm will not be received.
2). In a case when the text requisition is received real-time,the "STOP" designation is ignored and the text requisition
will be received in real-time at the start of the next analysis.
(2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "Continue",
the test requisition for the next sample will be received.
<Description>
(1). Communication error occurs when transmitting the text for test requisition or receiving the test requisition format.
(2). The display in the [ ] and its meaning are as follows.
aa : type of the errors
----------------------------------------------------------------
01 : Device name error
02 : Framing error
03 : Over run error
04 : Parity error
05 : Time out error
06 : Receive NAK when transmitting a text
07 : BCC error when receiving a text
08 : Other communication errors
09 : Function error (Full Duplex instead of Half Duplex)
10 : Unit name error
11 : Parameter error
12 : Request cancel
----------------------------------------------------------------
bbbbbbbbb : Text having error
----------------------------------------------------------------
RB : Start text for receiving the test
requisition information.
RE : End text for receiving the test
requisition information.
R_ __ nnnn : normal run, normal sample, sample No.
R_ _E nnnn : normal run, emergency sample, sample No.
R_ _P nnnn : normal run, STAT sample, sample No.
R_ U_ nnnn : normal run, urine sample, sample No.
R_ UE nnnn : normal run, urine emergency sample, sample No.
R_ UP nnnn : normal run, urine STAT sample, sample No.
R_ X_ nnnn : normal run, Other normal sample, sample No.
R_ XE nnnn : normal run, Other emergency sample, sample No.
R_ XP nnnn : normal run, Other STAT sample, sample No.
RH __ nnnn : repeat run, normal sample, sample No.
RH _E nnnn : repeat run, emergency sample, sample No.
RH _P nnnn : repeat run, STAT sample, sample No.
RH U_ nnnn : repeat run, urine sample, sample No.
RH UE nnnn : repeat run, urine emergency sample, sample No.
RH UP nnnn : repeat run, urine STAT sample, sample No.
RH X_ nnnn : repeat run, Other normal sample, sample No.
RH XE nnnn : repeat run, Other emergency sample, sample No.
RH XP nnnn : repeat run, Other STAT sample, sample No.
R_ __ xxxxxxxxxxx: normal run, normal sample, sample ID
R_ _E xxxxxxxxxxx: normal run, emergency sample, sample ID
R_ _P xxxxxxxxxxx: normal run, STAT sample, sample ID
R_ U_ xxxxxxxxxxx: normal run, urine sample, sample ID
R_ UE xxxxxxxxxxx: normal run, urine emergency sample, sample ID
R_ UP xxxxxxxxxxx: normal run, urine STAT sample, sample ID
R_ X_ xxxxxxxxxxx: normal run, Other normal sample, sample ID
R_ XE xxxxxxxxxxx: normal run, Other emergency sample, sample ID
R_ XP xxxxxxxxxxx: normal run, Other STAT sample, sample ID
RH __ xxxxxxxxxxx: repeat run, normal sample, sample ID
RH _E xxxxxxxxxxx: repeat run, emergency sample, sample ID