11)ONLINE ANALYSIS METHOD MISMATCH [ aa <> bb ]
<Operation>
(1).Disregard the test requisition information text for the sample generating the alarm.
(2).Continue to receive the test requisition information text for other samples within the run.
<Description>
(1).When receiving the test requisition information text using online, sample type of this text does not match
the required sample type.
(2).The display of [ ] and its meaning are as follows.
aa : Sample type
----------------------------------------------------------------
__ : Normal sample
_E : Emergency sample
_P : Stat sample
U_ : Urine normal sample
UE : Urine emergency sample
UP : Urine stat sample
X_ : Other normal sample
XE : Other emergency sample
XP : Other Stat sample
----------------------------------------------------------------
bb : Received sample type
----------------------------------------------------------------
__ : Normal sample
_E : Emergency sample
_P : Stat sample
U_ : Urine normal sample
UE : Urine emergency sample
UP : Urine stat sample
X_ : Other normal sample
XE : Other emergency sample
XP : Other Stat sample
----------------------------------------------------------------
12)ONLINE SAMPLE No. MISMATCH [ aaaaaa <> bbbbbb ]