Agilent E4408B Specifications and Characteristics
General
Agilent E4408B Specifications and
Characteristics
Specifications Supplemental Information
Power Requirements
ac Operation
Voltage, frequency 90 to 132 Vrms, 47 to 440 Hz
195 to 250 Vrms, 47 to 66 Hz
Power Consumption, On <300 W
Power Consumption,
Standby
<5 W
dc Operation
Voltage 12 to 20 Vdc
Power Consumption <200 W
Power Consumption,
Standby
<100 mW
Specifications Supplemental Information
Measurement Speed
Local Measurement and
Display Update rate
a
28/s, characteristic
Remote Measurement and
GPIB Transfer Rate
b,c
(Option A4H)
28/s, characteristic
RF Center Frequency Tune,
Measure, and GPIB Transfer
Time
b,d
(Option A4H)
90 ms, characteristic
a. Factory preset, auto align Off, fixed center frequency, RBW = 1 MHz, spans >10 MHz and
600 MHz, and stop frequency 3GHz.
b. Display Off (:DISPlay:ENABle OFF), and 32-bit integer data format (:FORMat:DATA
INT,32), if Option A4J is installed, disable sweep ramp, (:SYS-
Tem:PORTs:IFVSweep:ENABle OFF), markers Off, single sweep, measured with IBM
compatible PC with 550 MHz PentiumIII running Windows NT 4.0, one meter GPIB
cable, National Instruments PCI-GPIB card and NI-488.2 DLL
c. Factory preset, auto align Off, fixed center frequency, RBW = 1 MHz, and span = 20 MHz,
stop frequency 3 GHz, average of 100 measurements.
d. Factory preset, auto align Off, RBW = 1 MHz, span = 20 MHz, stop frequency 3GHz,
center frequency tune step size = 50 MHz.