Chapter 1 9
Agilent E4403B Specifications and Characteristics
About This Chapter
Agilent E4403B Specifications and
Characteristics
o If Auto Align Off is selected:
When the analyzer is at a constant temperature, within the
operating temperature range, for a minimum of 90 minutes.
After the analyzer is turned on for a minimum of 90 minutes, the
front panel amplitude reference has been connected to the
INPUT, and
Align Now All has been run.
When
Align Now All is run:
Every hour
If the ambient temperature changes more than 3 °C
If the 10 MHz reference changes
When Align Now RF is run (with the front-panel amplitude
reference connected to the INPUT):
Every 24 hours
If the ambient temperature changes more than 30 °C
o If
Auto Align All but RF is selected:
When the analyzer is at a constant temperature, within the
operating temperature range, for a minimum of 90 minutes.
After the analyzer is turned on for a minimum of 90 minutes, the
front panel amplitude reference has been connected to the
INPUT, and Align Now RF has been run.
When
Align Now RF is run (with the front-panel amplitude
reference connected to the INPUT):
Every hour
If the ambient temperature changes more than 3 °C