Agilent Technologies E4411B Water Dispenser User Manual


 
Chapter 3 83
Agilent E4411B Specifications and Characteristics
Amplitude
Agilent E4411B Specifications and
Characteristics
Specifications Supplemental Information
Display Range
Log Scale Ten divisions displayed;
0.1, 0.2, 0.5 dB/division and 1
to 20 dB/division in 1 dB steps
RBW 1 kHz Calibrated 0 to 85 dB from
Reference Level
RBW 300 Hz (Option 1DR)
Calibrated 0 to 120 dB
a
from
Reference Level
Linear Scale Ten divisions
Scale Units dBm, dBmV, dBµV, dBµA, A,
V, and W
a. 0 to 70 dB range when span = 0 Hz, or when IF Gain fixed:
(:DISPlay:WINDow:TRACe:Y[:SCALe]:LOG:RANGe:AUTO OFF).
Specifications Supplemental Information
Marker Readout
Resolution
Log scale
RBW 1 kHz
0 to 85 dB from ref level 0.04 dB
RBW 300 Hz (Option 1DR)
0 to 120 dB from ref level 0.04 dB
Linear scale 0.01% of Reference Level