Notes: Refer to the serial number plate on rear of the chiller for model and electrical data.
Cooling Capacity, Watts x 3.41 = BTUs/hr. Performance specifications determined at ambient temperature of 20°C/68°F.
For 50Hz models, derate cooling capacity and pump flow by 17%.
External Pressure Reducing Assembly (for Positive Displacement Pump models) Steps down high outlet pressure to 10 to 45 psi. Cat. No 060302
5
6000 Series Standard Chillers with Magnetic Drive Centrifugal Pump
Model 6205 6305 6505
Operating Temperature -5° to 40°C -5° to 40°C -10° to 40°C
Compressor
1
/
4 hp
1
/
3 hp
1
/
2 hp
Cooling Capacity @ 20°C 750 Watts 1200 Watts 1320 Watts
10°C 450 Watts 825 Watts 1055 Watts
0°C 100 Watts 375 Watts 550 Watts
Pressure at 0 Flow Rate 10 psi 10 psi 10 psi
Flow Rate @ 0 psi 4.3 gpm / 16.3 lpm 4.3 gpm / 16.3 lpm 4.3 gpm / 16.3 lpm
Reservoir Capacity 1.65 gal / 6.25 liters 1.65 gal / 6.25 liters 1.65 gal / 6.25 liters
Dimensions (in.) 24 x 14
3
/
8
x 26
1
/
2
24 x 14
3
/
8
x 26
1
/
2
24 x 14
3
/
8
x 26
1
/
2
h x w x d (cm) 61 x 36.5 x 67.3 61 x 36.5 x 67.3 61 x 36.5 x 67.3
Shipping Weight 155 lbs 165 lbs 170 lbs
Volts, Hz, Amps 120V, 60Hz, 6.8A 120V, 60Hz, 8A 120V, 60Hz, 13A
Volts Range 108V to 132V 108V to 132V 108V to 132V
60Hz Cat. No. 080501 080511 080521
Volts, Hz, Amps 240V, 50Hz, 3.4A 240V, 50Hz, 4A 240V, 50Hz, 6.5A
Volts Range 198V to 264V 198V to 264V 198V to 264V
Over Voltage: Category II
50Hz Cat. No. 080506 080516 080526
Additional Specifications for "P" Series 6000 Chillers with Positive Displacement Pump
Model "P" Series 6205P 6305P 6505P
Flow Rate @ 0 psi 1 gpm / 3.8 lpm 1 gpm / 3.8 lpm 1 gpm / 3.8 lpm
Pump Pressure (Adjustable) 20 to 100psi 20 to 100psi 20 to 100psi
Shipping Weight 160 lbs 170 lbs 170 lbs
Volts, Hz, Amps 120V, 60Hz, 9.6A 120V, 60Hz, 10.8A 120V, 60Hz, 15.8A
Volts Range 108V to 132V 108V to 132V 108V to 132V
60Hz Cat. No. 080502 080512 080522
Volts, Hz, Amps 240V, 50Hz, 4.8A 240V, 50Hz, 5.4A 240V, 50Hz, 8A
Volts Range 198V to 264V 198V to 264V 198V to 264V
Over Voltage: Category II
50Hz Cat. No. 080507 080517 080527
Additional Specifications for "T" Series 6000 Chillers with Turbine Pump
Model "T" Series 6205T 6305T 6505T
Flow Rate @ 0 PSI 3.5 gpm / 13.2 lpm 3.5 gpm / 13.2 lpm 3.5 gpm / 13.2 lpm
Pump Pressure (Adjustable) 20 to 100psi 20 to 100psi 20 to 100psi
Volts, Hz, Amps 120V, 60Hz, 9.6A 120V, 60Hz, 10.8A 120V, 60Hz, 15.8A
Volts Range 108V to 132V 108V to 132V 108V to 132V
60Hz Cat. No. 080503 080513 080523
Volts, Hz, Amps 240V, 50Hz, 4.8A 240V, 50Hz, 5.4A 240V, 50Hz, 8A
Volts Range 198V to 264V 198V to 264V 198V to 264V
Over Voltage: Category II
50Hz Cat. No. 080508 080518 080528
Environmental Conditions • Indoor Use Only
• Maximum Altitude : 2000 meter • Operating Ambient : 5° to 30°C
• Relative Humidity : 80% for temperatures to 30°C • Pollution Degree : 2
• Class 1 : Residential, Commercial, Light Industrial • Class 2 : Heavy Industrial