HP (Hewlett-Packard) Refrigerated Recirculating Chillers Refrigerator User Manual


 
Notes: Refer to the serial number plate on rear of the chiller for model and electrical data.
Cooling Capacity, Watts x 3.41 = BTUs/hr. Performance specifications determined at ambient temperature of 20°C/68°F.
For 50Hz models, derate cooling capacity and pump flow by 17%.
External Pressure Reducing Assembly (for Positive Displacement Pump models) Steps down high outlet pressure to 10 to 45 psi. Cat. No 060302
6
6000 Series Standard Chillers with Magnetic Drive Centrifugal Pump
Model 6705 6105 6155
Operating Temperature -15° to 40°C -15° to 40°C +5° to 35°C
Compressor
3
/
4 hp 1 hp 1
1
/
2
hp
Cooling Capacity @20°C 2500 Watts @20°C 2850 Watts @20°C 5200 Watts
@10°C 1700 Watts @10°C 1925 Watts @10°C 3600 Watts
@ 0°C 760 Watts @ 0°C 940 Watts @ 5°C 2700 Watts
Pressure at 0 Flow Rate 10 psi 10 psi 10 psi
Flow Rate @ 0 psi 4.3gpm / 16.3 lpm 4.3 gpm / 16.3 lpm 4.3 gpm / 16.3 lpm
Reservoir Capacity 1.65 gal / 6.25 liters 1.65 gal / 6.25 liters 6.25 gal / 23.65 liters
Dimensions (in.) 24 x 14
3
/
8
x 26
1
/
2
24 x 14
3
/
8
x 26
1
/
2
30 x 21x 27
h x w x d (cm) 61 x 36.5 x 67.3 61 x 36.5 x 67.3 76.2 x 53.3 x 68.6
Shipping Weight 178 lbs 178 lbs 450 lbs
Volts, Hz, Amps 208-230V, 60Hz, 7.5A 208-230V, 60Hz, 8A 208-230V, 60Hz, 12A
Volts Range 187V to 253V 187V to 253V 187V to 253V
60Hz Cat. No. 080531 080542 080593
Volts, Hz, Amps 240V, 50Hz, 7.5A 240V, 50Hz, 8.5A 240V, 50Hz, 12A
Volts Range 198V to 264V 198V to 264V 198V to 264V
Over Voltage: Category II
50Hz Cat. No. 080536 080546 080594
Additional Specifications for "P" Series 6000 Chillers with Positive Displacement Pump
Model "P" Series 6705P 6105P 6155P
Flow Rate @ 0 psi 4 gpm / 15.2 lpm 4 gpm / 15.2 lpm 4 gpm / 15.2 lpm
Pump Pressure (Adjustable) 20 to 100psi 20 to 100psi 20 to 100psi
Shipping Weight 178 lbs 178 lbs 450 lbs
Volts, Hz, Amps 208-230V, 60Hz, 10A 208-230V, 60Hz, 11A 208-230V, 60Hz, 12A
Volts Range 187V to 253V 187V to 253V 187V to 253V
60Hz Cat. No. 080532 080544 080596
Volts, Hz, Amps 240V, 50Hz, 10A 240V, 50Hz, 11A 240V, 50Hz, 12A
Volts Range 198V to 264V 198V to 264V 198V to 264V
Over Voltage: Category II
50Hz Cat. No. 080537 080547 080597
Additional Specifications for "T" Series 6000 Chillers with Turbine Pump
Model "T" Series 6705T 6105T 6155T
Flow Rate @ 0 PSI 3.5 gpm / 13.2 lpm 3.5 gpm / 13.2 lpm 3.5 gpm / 13.2 lpm
Pump Pressure (Adjustable) 20 to 100psi 20 to 100psi 20 to 100psi
Volts, Hz, Amps 208-230V, 60Hz, 9.6A 208-230V, 60Hz, 10.8A 208-230V, 60Hz, 12A
Volts Range 108V to 132V 108V to 132V 187V to 253V
60Hz Cat. No. 080533 080543 080598
Volts, Hz, Amps 240V, 50Hz, 9.6A 240V, 50Hz, 10.8A 240V, 50Hz, 12A
Volts Range 198V to 264V 198V to 264V 198V to 264V
Over Voltage: Category II
50Hz Cat. No. 080538 080548 080599
Environmental Conditions • Indoor Use Only
• Maximum Altitude : 2000 meter • Operating Ambient : 5° to 30°C
• Relative Humidity : 80% for temperatures to 30°C • Pollution Degree : 2
• Class 1 : Residential, Commercial, Light Industrial • Class 2 : Heavy Industrial