RTI ID900 Refrigerator User Manual


 
12
VI. Specifications
Sample Parameters:
Vapor only, oil and liquid free, 300 psig (2 MPa) Maximum
Detected Compounds:
R12, R134a, Air and unknown compounds
Sensor Technologies:
Non-Dispersive Infrared and Chemical Fuel Cell
Refrigerant Sample Size:
0.12 ounces (3.5 grams) per typical sample – weight of vapor loss
Power (Standard):
Powered by vehicle battery through clip-on adapter harness (10 to 14 VDC).
Power (Optional):
Powered by line power through plug-in wall adapter harness, 115 VAC unit (OPTIONAL) or 220 VAC unit (OPTIONAL)
Operational Temperature Range:
45-100
o
F (7-38
o
C)
Instrument Weight:
1.5 Lbs. (454 grams)
R12/R134a Purity Setpoint:
Pass if single refrigerant, R12 or R134a, is detected in concentrations equal to or greater than 98% by weight.
Unknown Refrigerant Purity Setpoint:
In the presence of detected unknown refrigerant types R12 or R134a is considered pure if detected in concentrations equal to or
greater than 96% by weight.
Effect of Air Concentrations:
Air is not considered a contaminate and will be removed from the weight concentration determination of refrigerants. Air
concentrations will only be reported for blends of R12 and R134a refrigerants.
SAE J1771:
If the refrigerant being tested is identified as contaminated, any visual percentages displayed of CFC-12 (R-12) or HFC-134a (R-
134a) outside the design certified value is informational and may not be accurate. SAE J1771 Certification Note: This equipment
has not been designed certified for contaminate detection of R-124 or R-142B.