Sanyo MPR-414F Refrigerator User Manual


 
54
22
OPTIONAL COMPONENT
Installation of recorder
Refrigerator (MTR-0620LH)
1. Attach the fixture to the temperature recorder with
two screws. (Fig. 1)
2. Remove the screws at the right corner on the top
of the unit and fix the recorder to the attachment
position. (Fig. 2)
3. Remove a rubber cap and insulation covering the
access port on the back of the unit.
4. Take the recorder sensor into the refrigerator
chamber through the access port.
5. Fix the recorder sensor on the right wall of the
chamber with the enclosed clips. (Fig. 3)
6. Arrange the capillary tube so that it cannot disturb
the shelves. Take care not to crush the tube when
it is bent.
7. Make a cut on the rubber cap and pass the
capillary tube through it. (Fig. 4)
8. Cover the access port with the insulation and
rubber cap.
Note:
The temperature recorder MTR-0620LH can be installed also at the mounting space of temperature
recorder (bottom right of the cabinet) if the recorder for freezer (MTR-4014LH) is not installed. In this
case, a dedicated recorder fixture (MPR-S30) is needed.
For the installation, refer to “For freezer (MTR-4014LH)” below.
Fig. 1
Fig. 2
Fig. 3
Fig. 4