MIN/MAXDataRecording(REC)mode
Thisinstrumenthastheabilityrecordthehighest(MAX)andthelowest(MIN)readings
recorded.ReadingsareretainedevenaftertheinstrumentisswitchedOFF.
1. MomentarilypresstheRECbutton(RECwillbedisplayedattheupperright).
2. PressRECagain.REC/MAXwillbedisplayedalongwiththemaximum
reading.
3. PressRECagaintodisplaythe MINreading.REC/MINwilldisplayalongwiththeminimum
reading.
4. TodeleteeithertheMAXortheMINreading,presstheHOLDbuttonwhileeitherthe
REC/MAXortheREC/MINiconsaredisplayed.
5. PressandholdRECfor2seconds
toexitandtoreturntothenormaloperatingmode.The
RECandMAX/MINiconswillswitchOFF.
Specifications
Measurements/Parameters Range(Resolution)
Accuracy
SolutionTemperaturemeasurement 0to100
o
C(32to212
o
F)
0.8
o
C(±1.5
o
F)
pHmeasurement
0.00to14.00pH(0.01pH)
±0.04pH
mV(ORP)measurement
±1999mV
±0.5mV
Note:AccuracydoesnotincludepHbuffertemperaturecoefficient
AutomaticTemperature
Compensation(ATC)range
0to65
o
C(32to149
o
F)
ManualTemperatureCompensation
Adjustmentrange
0to100
o
C(32to212
o
F)
Power 9voltalkalinebattery
Display LCDdisplayforsimultaneouspHormVand
temperaturereadingsplusprogrammingmenu
parametersandicons
OperatingTemperatureandRelative
Humidity
0to50
o
C(32to122
o
F);<80%RH
pHCalibrationpoints 4.00,7.00,10.00(1,2,or3pointcalibrationcanbe
performed)
AutoPowerOFF 10minutesfromlastkey‐press(canbedisabled)
LowBatteryIndication Batterydisplayiconwillflashwhenbatteryrequires
replacing
Dimensions/Weight 135x60x33mm(5.3x2.4x1.3”);200g(7.1oz.)