BatteryReplacement
The9Vbatterythatpowerstheinstrumentrequiresreplacingwhenthebatteryiconflasheson
thedisplay.Toreplacethebatteryfollowthestepsbelow:
1. Disconnecttheelectrode/probefromtheinstrument.
2. OpentherearbatterycompartmentusingascrewdrivertoremovethetwoPhillipshead
screwsthatsecurethecompartment.
3. Carefullyremovetheoldbatteryfromthecompartmentlidchamberandgentlyunsnap
thebattery’sterminalsfromthewired(red/black)connectors.
4. Installanew9Vbatterybysnappingthebatteryterminalsontothewiredconnectors
(observingcorrectpolarity)andinsertingthebatteryintothecompartmentlidchamber.
5. Installthecompartmentlidchamberintothemeterhousingandsecurewiththetwo
screws.
Neverdisposeofusedbatteriesorrechargeablebatteriesinhouseholdwaste.
Asconsumers,usersarelegallyrequiredtotakeusedbatteriestoappropriate
collectionsites,theretailstorewhere thebatterieswerepurchased,orwherever
batteriesaresold.
Disposal:Donotdisposeofthisinstrumentinhouseholdwaste.Theuserisobligatedtotake
end‐of‐lifedevicestoadesignatedcollectionpointforthedisposalofelectricalandelectronic
equipment.
ElectrodeMaintenanceandTroubleshooting
Mostmeasurementdifficultiesarecausedbytheelectrode.Carefulhandlingandproperstorageof
theelectrodewillincreasemeasurementaccuracyandelectrodelife.
Electrodesshouldberinsedoftentoremovealltracesofstoragesolution,processmedium,or
previoustestsolution.
ElectrodeStorage
TheelectrodeshouldbestoredinasoakerbottleorwettingcapfilledwithpH4buffer.Never
storeanelectrodeindistilledordeionizedwaterasthiswillcausemigrationofthefillsolution
fromtheelectrode.
ElectrodeCleaning
Thetypeofelectrodecoatingthatoccursovertimewilldeterminethecleaningmethod.Soft
coatings can be removed by vigorous stirring or by using a squirt bottle. Organic chemical
coating(hard coatings)shouldbe chemicallyremoved.Onlyinextremecasesshouldthebulbbe
cleanedmechanically.