PH210-en-GB_v1.3 11/14
8
TroubleshootinganElectrode
Symptom Cause Recommendation
Longresponsetimeor
readingdrift
CloggedJunction Soakin4.07MPotassiumChloride(KCl)
@60
o
Cfor30minutes.
StrongAlkaline
measurement
Soakin0.1MHCLovernight.
DeterioratedGellayer Replaceelectrode.
Proteincoatingonelectrode
surface
Soakin1gmPepsindissolvedin100ml
of0.1mHCLfor30minutesoras
needed.
Oil,paint,dyes,suspended
solidsetc.areonsensor
Rinseelectrodealternatelywitha
materialssolventandabuffer7.00.
Organicsolventsarecoating
thesensor
Organicmolefractionhastobeless
than50%toassurereasonablereadings.
Limittimeofmeasurement.Keepprobe
inthe7.0bufferbetweenreadings.
Dehydratedmembrane Readsolutionfordrybulbbelow:
DryBulb Longtermstoragewithout
wetting
Soakelectrodetipinwettingcapfilled
with1ml7.00bufferfor24to48hours
StaticCharge Wipingelectrodes Rinseelectrodein7.0bufferandblot.
Donowipeelectrode.
Samereadingsin
differentbuffersand
samples
Crackedorbrokenbulb Replaceelectrode.Usebulbguard.
Avoidplungingelectrodetobottomof
containerandspinningbars.Wetting
capwillprotectbulbbetweentests.
ErraticLCDdisplay Sampleshavelowionic
strength(lackssalt);e.g.
distilled,deionized,boiled,
lakewater(highpressure)
Foreach50mlofsampleadd1drop
(50uL)ofSAT.KCl(NoalterationinpH
willoccurusinginertKCl).