Blood/Fluid and Irrigation Warming Systems
English
22
Check the 3M
TM
Ranger
TM
system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A
Specifications
phySiCAl ChARACteRiStiCS
Warming unit
4.5 in. (11 cm) high x 7.5 in. (19 cm) wide x 10 in. (25 cm) long; wt.: 7 lb. 7 oz. (3.4 kg)
certiFicationS
IEC/EN 60601-1; CAN/CSA-C22.2, No. 601.1
claSSiFication
MEDICAL — GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE AND
MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1; CAN/CSA-C22.2,
No.601.1; ANSI/AAMI ES60601-1:2005 CSA-C22.2 No. 60601-1:08; Control No. 4HZ8
Classied under IEC 60601-1 Guidelines (and other national versions of the Guidelines) as Class I,
Type B, Ordinary equipment, Continuous operation. Classied by Underwriters Laboratories Inc. with
respect to electric shock, re and mechanical hazards only, in accordance with IEC/EN
2
60601-1 and in
accordance with Canadian/CSA C22.2, No. 601.1. Classied under the Medical Device Directive as a
Class IIb device.
leAkAge CURRent
Meets leakage current requirements in accordance with UL/IEC 60601-1.
enviRonmentAl ConditionS
operating temperature range
15° to 40°C (59° to 104°F)
Storage temperature range
-20° to 45°C (-4° to 113°F)
operating humidity
10 to 85% RH, noncondensing
atmoSpheric preSSure range
50 kPa to 106 kPa
eleCtRiCAl ChARACteRiStiCS tempeRAtURe ChARACteRiStiCS
device rating
100-120 VAC, 50/60 Hz
220-240 VAC, 50/60 Hz
Set point temperature
41°C
maximum heating poWer
900 W
over-temperature alert
Model 245:
43°C
Model 247:
48°C
FuSe
2 x T10A-H (250V) for 100-120 VAC
2 x T6.3A-H (250V) for 220-240 VAC
under-temperature alert
33°C
FuSe type
Time delay, high breaking
over-temperature cutoFF
Model 245:
44°C (warming units Rev N and newer)
46°C (warming units Rev A to M)
Model 247:
50°C