OPTIONAL EQUIPMENT
11.1 COBEX CHART RECORDER INSTALLATION: TFO-10 & TFO-28
The following installation procedure is for installing a 6” Cobex Chart Recorder in a Cascade TEK TFO-10
or TFO-28 Forced Air Oven:
Uncrate oven. Do not connect power. Locate Chart recorder cut-out at the right corner of the instrument
panel. Remove cut out cover. Remove the top of the oven to gain access to the electrical compartment.
Insure the chart recorder body has the “L” shaped mounting brackets. Insert the chart recorder into the
recorder opening.
Using the “L” shaped brackets located on the top and sides of the recorder body, adjust and tighten the
brackets against the inside wall of the instrument panel.
Locate the temperature probe hole in the top of the oven. Located right behind the back of the recorder
allowing access to the oven work space.
Feed the temperature probe into the oven workspace. Attach the probe to the bracket on the right side
wall of the oven.
Note: The bracket will have the temperature probe from the oven controller already in place.
Located the Red and Black wires coming from the oven wiring harness and labeled “RECORDER”. From
the recorder, attach the black and white power wires to those leads with a wire cap. Attached the green or
ground wire to any oven body screw that will provide good metal to metal contact.
Refer to the Chart Recorder Manual for operation and calibration.